Installing/Mounting

4.3 MAG 5000/6000 compact

4.3
MAG 5000/6000 compact
Install MAG 5000 / MAG 6000 compact version
22
www.eltra-trade.com
1. Remove and discard terminal box lid of sensor.
2. Ensure SENSORPROM
3. Fit M20 or ½" NPT cable glands for supply and output cables.
4. Unplug the two black plug assemblies for coil and electrode cables in terminal box.
5. Connect earth wire from connection board to bottom of terminal box.
6. Connect 2-pin connector and 3-pin connector as shown to their corresponding terminal
numbers on connection board as shown in Electrical connection (Page 36).
Note
System will not register flow if black plugs are not connected to connection board.
7. Fit supply and output cables through cable glands and connect to connection plate as
shown in Electrical connection (Page 36).
memory unit is installed.
®