Power connection
Only use the proprietary power supply. Make a proper connection by ensuring that the shrink tubing
completely secures the connection between the DC power cord and the extension cord.
Electrical input rating: 13 VDC, 6.92A
Only use a hospital grade power supply cord.
Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle
labeled "Hospital Only" or "Hospital Grade".
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
The equipment is intended for continuous operation.
Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Water and moisture
Never expose the display to rain or moisture.
Degree of protection against the ingress of water: IP23 compliance.
Ventilation
Do not cover the slots or openings of the monitor for proper heat dissipation. Always put the monitor in a
place where there is adequate ventilation to prevent internal heat buildup.
Do not block the monitor cooling vents. The monitor is cooled by natural convection and has no fan
Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may
block the ventilation slots.
Warnings
1. Federal law (United States of America) restricts this device to use by, or on order of a physician.
2. Should any solid object or liquid fall into the panel, unplug the unit and have it checked by qualified
personnel before operating it any further.
3. Unplug the unit if it is not to be used for an extended period of time. To disconnect the cord, pull it out by
the plug. Never pull the cord itself.
4. Test this equipment prior to a medical procedure. This monitor was fully tested at the factory before
shipment.
5. Do not attempt internal repairs or adjustments not specifically detailed in this operating manual.
6. Pay close attention to the care, cleaning instructions in this manual. A deviation may cause damage.
7. Do not sterilize the monitor.
8. Read the entire operating manual before assembling or connecting the camera.
9. Do not stack more than 3 boxes high.
10. The device has no means to be incorporated in an IT-network in the clinical environment.
11. The enclosure has to be checked upon collision damage, refer to qualified service personnel.
12. The protective screen is made of tested high-resistance PMMA (polymethyl methacrylate). Nonetheless
there is the possibility that it may crack if subject to strong impacts. Evaluate and prevent the risk of
possible breakages of the protective screen by correctly handling and positioning the monitor in the
operating room.
Cautions
1. Turn the power off when the unit is not in use.
2. Never operate the unit right after having transported from a cold location directly to a warm location.
3. Do not force the monitor past 28 degrees of vertical when adjusting the screen position (for monitors
equipped with stands only).
4. Remove the power module and connection when transporting the unit.
5. Do not install the unit near mechanical vibration.
6. The unit is designed for operation in a horizontal position. Never operate the unit in a vertical position.
7. Keep the unit away from equipment with strong magnets (i.e. a large loudspeaker).
8. The products have lower breaking capacity type. Do not install at the building power system prospective
short-circuit current exceeding 35 A.
Important information
R5912860 /13
AMM215WTD
33