Specifications And Environmental Conditions; Specifications - Fluke 9102S User Manual

Dry-well calibrator
Hide thumbs Also See for 9102S:
Table of Contents

Advertisement

Specifications and Environmental Conditions

Specifications

Range
Accuracy
Stability
Resolution
Well-to-Well Uniformity
Heating Times
Stabilization
Cooling Times
Well Depth
Removable Inserts
Power Requirements
Power Rates
Size
Environmental Conditions
Operation Temperature Range
Ambient Relative Humidity
Ambient Pressure
Altitude
Weight
Safety
Electromagnetic Compatibility (EMC)
3
Table 1. Specifications
-10 °C to 122 °C (14 °F to 252 °F) at an ambient of 23 °C
±0.25 °C
±0.05 °C
0.1 °C or °F
±0.2 °C with sensors of similar size at equal depths within wells
ambient to 100 °C: 10 minutes
7 minutes
ambient to 0 °C: 10 minutes
102 mm (4 inches)
See
Accessories
100 to 240 V ac, 50/60 Hz, 90 VA; or 12 V dc, 2.6 A
100 mm H x 152 mm W x 175 mm D (4 in x 6 in x 6.9 in)
5 to 50 °C (41 to 122 °F)
0 to 80 % RH for temperatures up to 30 ºC, 0 to 50 % RH for temperatures 30 to 50 °C
75 kPa to 106 kPa
<2000 m
4 lb (1.8 kg)
Mains
IEC 61010-1: Overvoltage Category II, Pollution Degree 2
Additional
IEC 61010-2-010 (Heating)
International................................. IEC 61326-1: Basic Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those
directly connected to a low-voltage power supply network that supplies buildings used for
domestic purposes. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is
connected to a test object.
The equipment may not meet the immunity requirements of this standard when test leads
and/or test probes are connected.
Korea (KCC) ............................... Class A Equipment (Industrial, Broadcasting, &
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and
the seller or user should take notice of it. This equipment is intended for use in business
environments and not to be used in homes.
USA (FCC).................................. 47 CFR 15 subpart B. This product is considered an exempt
Specifications and Environmental Conditions
Communication Equipment)
device per clause 15.103.
Dry-well Calibrator

Advertisement

Table of Contents
loading

Table of Contents