Autotest in Far End Source Mode
1
Turn on the tester and smart remote and let them sit for
5 minutes. Allow additional time if the modules have
been stored above or below room temperature. For
other sources, warm up according to the manufacturer's
recommendations.
2
Turn the rotary switch to SETUP, then select Fiber. Set
the following on the Fiber tabs (press
tabs):
Fiber Type: Select the fiber type to be tested.
Test Limit: Select the test limit required for the job.
J
Press
More to see other lists of limits.
Remote End Setup: Set to Far End Source
Bi-directional: Does not apply to Far End Source
mode.
C
to see other
See Table 2 on page 12 for details on settings.
Number of Adapters and Number of Splices: Enter
the number of adapters and splices that will be
added to each direction of the fiber path after the
reference is set.
Connector Type: Select the connector type used in
the cabling to be tested. Select General if the exact
type is not listed.
Test Method: Refers to the number of adapters
represented in the loss results. Select Method B if
you use the reference and test connections shown
in this manual.
Certifying Fiber Cabling
-continued-
29