What to do if ...
Problem
(flashing)
e.g.
................
e.g.
................
Measurement
or inflation
does not start
instantly.
Readings from
repeated mea-
surements differ
considerably.
Prior to a
measurement,
the positioning
system (A) on the
display does not
work.
12
Reason
Batteries are flat.
Memory readings can be
called up.
Measurements are not
possible.
Error codes 01-29:
Device may not be
correctly positioned on
the wrist.
Error codes 30 and
higher:
System error
Right measuring position
was not reached.
Blood pressure is a
fluctuating value.
For healthy adults,
deviations of about 10 to
20 mmHg are possible.
Cuff is not properly fitted. Ensure that the appliance is fitted
Readings were not con-
sistently taken at heart
level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
Positioning system (A) is
deactivated.
Solution
Insert new batteries.
Refit device and repeat
measurement as shown and
described under «Measuring».
Repeat measurement. If problem
persists, have device checked at a
Braun Service Centre.
Repeat measurement as shown
and described under «Measuring»
or deactivate the positioning
system (A).
none
as follows:
– on top of the slim side of the
forearm
– with the display on top
– with the battery compartment
facing the user's body
– approx. 2 cm away from the
thumb root
For each measurement, proceed
exactly as shown and described
under «Measuring».
While measuring, relax, keep still,
do not move, or speak.
Refer to the chapter «Select mode»
to activate the positioning system
(A).