Troubleshooting; Specifications - Braun ExactFit 5 Connect Manual

Hide thumbs Also See for ExactFit 5 Connect:
Table of Contents

Advertisement

Available languages

Available languages

Troubleshooting

Problem
Reason
Heart rate
Appears in the measurement
symbol
condition and flashes when
pulse is detected
Low battery
Appears when the battery
indicator
voltage is excessively low or the
positions of the batteries are
incorrect
Measuring
Appears when the accurate
error
blood pressure and pulse could
not be obtained
E0
Measuring error
E1
Pumping error
E2
Out of Range Results Error
E3
Pump time error
E254
Device internal error
E255
Calibration error
E256
An error occurred during the
Bluetooth Firmware update
10
Solution
• Measurement in progress, remain quiet.
• Replace all four batteries with new ones.
Insert the batteries in the correct
positions. Be aware of +/- positions.
• Press "start/stop" button again and
remeasure.
• Check if cuff is wrapped according
to instructions.
• Check that there are no kinks in tube.
• Check that hand is relaxed.
• Check if talking or moving
during measurement.
• Check if posture is correct.
• The pulse signals on the cuff are too
weak. Re-position the cuff and repeat the
measurement.
• The cuff is not fastened properly. Re-position
the cuff and repeat the measurement.
• Relax for a while and repeat the
measurement. Make sure the cuff is correctly
positioned on your arm.
• The cuff is not fastened properly. Re-fit the
cuff, and take a measurement again.
• Remove and then reinsert the batteries. If the
problem persists, contact customer support.
• Remove and then reinsert the batteries. If the
problem persists, contact customer support .
• Device will automatically restart using the
original Bluetooth firmware. Re-initiate the
firmware download process or follow the on-
screen instruction from Braun APP.

Specifications

Method of measurement
Oscillometric
Model number
BUA6350
Range of measurement
Rated cuff pressure: 0mmHg~300mmHg
Measurement pressure: SYS: 60~255mmHg DIA: 40~200mmHg
Pulse value: 40-199 beat/minute
Accuracy
Pressure +/- 3 mmHg
Pulse +/- 5 % Max.
Inflation
Deluxe automatic
Display
LCD with backlight
Sets of device memory
60 sets per user
Sets of Bluetooth memory
200 sets per user
Cuff size
Small cuff = 22-32cm arm circumference
Large cuff = 32-42cm arm circumference
Operating condition
+10 °C ~ +40 °C, less than 85 % R.H.
Storage condition
-20 °C ~ +60 °C, less than 85 % R.H.
Operating atmospheric pressure
860-1060 hPa
Unit weight
Approximately 500gr (without batteries)
Power supply
Alkaline battery: 4 x AA (LR6) 1.5V
Battery life
300 times measurement
Auto power off
Whenever not used for 1 minute
Accessories
4 batteries, 2 arm cuffs with tube, instruction manual,
quick reference guide, storage pouch
Wireless communication
Bluetooth® Smart
Frequency Range: 2.4GHz ISM Band (2400 - 2483.5 MHz)
Modulation: GFSK
Effective radiated power: <20dBm
Service life:
5 years.
Important
Read the operating instructions.
If the device is not used within the specified temperature, humidity and atmospheric
pressure ranges, the technical accuracy of the measurement cannot be guaranteed.
Classification:
Internally powered equipment
Type BF applied parts
Protected against solid foreign objects of 12.5 mm diameter and larger.
Protected against water drops falling vertically when the device is tilted up to 15°.
Not suitable for use in the presence of flammable anaesthetic mixtures with air, oxygen or nitrous oxide.
Continuous operation with short-time loading.
40 °C
Operating temperature
10 °C
Storage humidity/operating humidity
Subject to change without notice.
English
Storage temperature
11