Depth Adjustment
7.20.1 Depth Adjustment
Observe the angle of the footplate for reinstallation.
The settings for positioning the footplates on the half-clamps
may vary for each footplate.
Remove the two flat screws A, washers B and locknuts C that
1.
secure footplate D to the half clamp E.
2.
Move footplate to one of four mounting positions.
If desired depth is still not obtained, rotate the half
clamp 180° on the footplate hinge F.
3.
Retighten the two flat screws, washers and locknuts.
1143192-L-03
7.20.2 Angle Adjustment
Observe the angle of the footplate for reinstallation.
The settings for positioning the footplates on the half-clamps
may vary for each footplate.
Loosen, but DO NOT remove, the two flat screws A, washer B
1.
and locknuts C that secure the footplate D to the half clamp E.
2.
Position the footplate to the necessary angle to accommodate
the user.
3.
Retighten the two flat screws, washers and locknuts.
7.20.3 Perpendicular and/or Inversion/Eversion
Adjustment
It is not necessary to remove the footplate to perform this
adjustment.
Insert a flathead screwdriver through the half clamp E on the
1.
footplate.
2.
Slowly turn nylon adjustment screw G in or out until footplate is
perpendicular to the footrest assembly or the desired inversion
or eversion is obtained.
Setup Maintenance
45