Features And Connections; Dc Current Measurement - Fluke i410 Instructions Manual

Ac/dc current clamp
Hide thumbs Also See for i410:

Advertisement

Features and Connections

See Table 1 for a list of features and connections.
1
[1] Minimum voltmeter requirements:
• Accepts safety-shrouded banana plugs
• Can display 1 mV (0.1 mV preferred)
• Accuracy ≥0.75 %
• Input impedance 1 MΩ, ≤100 pF

DC Current Measurement

See Table 2 for instructions on how to make a dc current measurement.
Maximum 400 A dc rms (i410) or 1000 A dc rms (i1010).
CAT
600V
600A
A
ZERO
i410
CLAMP METER
600V
Table 1. Features and Connections
2
8
CAT
600V
600A
3
4
5
A
ZERO
i410
CLAMP METER
600V
6
76
TRUE RMS MULTIMETER
9
V
7
Table 2. DC Current Measurement
CAT
600V
600A
A
ZERO
i410
CLAMP METER
600V
5
6
76
TRUE RMS MULTIMETER
3
mV
4
1
Item
Note
Step
2
Description
Jaw Lever
Jaw Centering Marks
ON Indicator
ON/OFF Switch
Zero Adjust
Battery Access
Output Cable
Tactile Barrier
[1]
Voltmeter
Description
Connect to voltmeter.
Select mV dc.
Set to ON.
Adjust ZERO (jaws empty).
Clamp and center around the
conductor.
Read voltmeter
(1 mV = 1 A).