Figure 3-7. Input Connections for 4-Wire Ohms Using 2x4 Wire Leads

Measuring Current

To avoid blowing the current fuse or damaging the Meter, do
not apply power to the circuit under test until test leads are
properly installed to the appropriate input terminals. For current
measurements above 200 mA, install the test leads into the 10 A
and LO terminals only.
The Meter is capable of making ac and dc current measurements up to 10 A.
To perform a current measurement:
1. Turn off power to the circuit under test.
2. Connect test leads between the Meter and the circuit under test.
3. If the circuit current is unknown, start by using the 10 A and LO terminals.
4. If the measurement is expected to be below 200 mA, connect the test leads to the
200 mA and LO terminals only and remove any leads in the 10 A terminal. See
Figure 3-8.
5. For measurements expected to be 200 mA to 10 A, connect the test leads to the
10 A and LO terminals only. See Figure 3-9.
6. Press B to measure ac current or press E to measure dc current.
7. Apply power to the circuit under test.
The Meter selects the appropriate range in autorange mode. The function and
measurement are displayed.
Operating the Meter from the Front Panel
Ground tab
W Caution
Selecting a Measurement Function
INPUT
SENSE
V
4 W
2W/4W
HI
HI
1000 V
CAT I
1000V
300V
MAX
600 V
CAT II
750V
1V
LO
LO
500 V pk
200 mA
10 A
MAX
MAX
mA
10A
eue26.eps
3
3-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

8808a

Table of Contents