1535/1537
Calibration Manual
Output Voltage 2.5 M
VAC 55 Hz
VAC 45 Hz
VAC 150 Hz
VAC 500 Hz
VDC
VDC
VDC
VDC
VDC
OHM
OHM
OHM
OHM
OHM
OHM
OHM
OHM
20
Table 4. Performance Test Points (cont.)
Function
Ω
Load
2500
5
600
600
600
-600
-10
0
300
600
0
10
30
500
1000
5000
10
60
Test Point
Value
Unit
V
V
V
V
V
V
V
V
V
V
Ω
Ω
Ω
Ω
Ω
Ω
Ω
k
Ω
k
Display Reading
Lower Limit
Upper Limit
2500
2650
3.9
6.1
587
613
587
613
587
613
-613
-587
-11.2
-8.8
-1
1
293
307
587
613
-1
1
8.8
11.2
28.4
31.6
489
511
970
1030
4890
5110
9.7
10.3
58.7
61.3