Fluke 1535 Calibration Manual page 16

Hide thumbs Also See for 1535:
Table of Contents

Advertisement

1535/1537
Calibration Manual
adc i r
cal_param current 7
-19 ADC_READ
irs 1
adc i r
cal_param current 8
-25 ADC_READ
adc i r
cal_param current 9
-190 ADC_READ
irs 0
adc i r
cal_param current 10
-250 ADC_READ
adc i r
cal_param current 11
-400 ADC_READ
cal stop_current
cal_param save
12
Read the Current analog-to-digital conversion value
Write the calibration parameter 7, use read value
ADC_READ in the previous step
Enter 200 μA range
Current source outputs -25 μA
Read the Current analog-to-digital conversion value
Write the calibration parameter 8, use read value
ADC_READ in the previous step
Current source outputs -190 μA
Read the Current analog-to-digital conversion value
Write the calibration parameter 9, use read value
ADC_READ in the previous step
Enter 2 mA range
Current source outputs -250 μA
Read the Current analog-to-digital conversion value
Write the calibration parameter 10, use read value
ADC_READ in the previous step
Current source outputs -400 μA
Read the Current analog-to-digital conversion value
Write the calibration parameter 11, use read value
ADC_READ in the previous step
Stop the Current calibration
Save the calibration parameters
ADC_READ "current_adc"
"current param"
"irs"
ADC_READ "current_adc"
"current param"
ADC_READ "current_adc"
"current param"
"irs"
ADC_READ "current_adc"
"current param"
ADC_READ "current_adc"
"current param"
"cal stop_current"
"params saved"