1535/1537
Calibration Manual
adc hv r
cal_param pwm 6
76.5 V_READ ADC_READ
pwm 78.2
adc hv r
cal_param pwm 7
78.2 V_READ ADC_READ
pwm 79.9
adc hv r
cal_param pwm 8
79.9 V_READ ADC_READ
pwm 81.5
adc hv r
cal_param pwm 9
81.5 V_READ ADC_READ
pwm 83.1
adc hv r
cal_param pwm 10
83.1 V_READ ADC_READ
pwm 84.7
adc hv r
cal_param pwm 11
84.7 V_READ ADC_READ
pwm 86.2
adc hv r
cal_param pwm 12
86.2 V_READ ADC_READ
cal stop_pwm
cal_param save
14
Read the HV analog-to-digital conversion value
Write the calibration parameter 6, use measured
value V_READ and read value ADC_READ in the
previous two steps
Set PWM ratio to 78.2 %
Measure the L-G voltage by the high voltage
meter, the measured (in Volts) is V_READ
Read the HV analog-to-digital conversion value
Write the calibration parameter 7, use measured
value V_READ and read value ADC_READ in the
previous two steps
Set PWM ratio to 79.9 %
Measure the L-G voltage by the high voltage
meter, the measured (in Volts) is V_READ
Read the HV analog-to-digital conversion value
Write the calibration parameter 8, use measured
value V_READ and read value ADC_READ in the
previous two steps
Set PWM ratio to 81.5 %
Measure the L-G voltage by the high voltage
meter, the measured (in Volts) is V_READ
Read the HV analog-to-digital conversion value
Write the calibration parameter 9, use measured
value V_READ and read value ADC_READ in the
previous two steps
Set PWM ratio to 83.1 %
Measure the L-G voltage by the high voltage
meter, the measured (in Volts) is V_READ
Read the HV analog-to-digital conversion value
Write the calibration parameter 10, use measured
value V_READ and read value ADC_READ in the
previous two steps
Set PWM ratio to 84.7 %
Measure the L-G voltage by the high voltage
meter, the measured (in Volts) is V_READ
Read the HV analog-to-digital conversion value
Write the calibration parameter 11, use measured
value V_READ and read value ADC_READ in the
previous two steps
Set PWM ratio to 86.2 %
Measure the L-G voltage by the high voltage
meter, the measured (in Volts) is V_READ
Read the HV analog-to-digital conversion value
Write the calibration parameter 12, use measured
value V_READ and read value ADC_READ in the
previous two steps
Stop the PWM calibration
Save the calibration parameters
ADC_READ "pwm_volt_adc"
"pwm param"
"pwm adjusted"
ADC_READ "pwm_volt_adc"
"pwm param"
"pwm adjusted"
ADC_READ "pwm_volt_adc"
"pwm param"
"pwm adjusted"
ADC_READ "pwm_volt_adc"
"pwm param"
"pwm adjusted"
ADC_READ "pwm_volt_adc"
"pwm param"
"pwm adjusted"
ADC_READ "pwm_volt_adc"
"pwm param"
"pwm adjusted"
ADC_READ "pwm_volt_adc"
"pwm param"
"cal stop_pwm"
"params saved"