8.
To calibrate the Ohm function (1537 only), use the PC and serial communication software to enter these
commands:
Command
Operation / Purpose
Connect the calibrator between E and G (use Ohm
function)
cal start_dmm_ohm
Enter OHM calibration mode
drs 1
Enter 600
Calibrator outputs 0
adc vr_vc r
Read the OHM analog-to-digital conversion value
cal_param dmm_ohm 0
Write the calibration parameter 0, use read value
0 ADC_READ
ADC_READ in the previous step
Calibrator outputs 600
adc vr_vc r
Read the OHM analog-to-digital conversion value
cal_param dmm_ohm 1
Write the calibration parameter 1, use read value
600 ADC_READ
ADC_READ in the previous step
drs 2
Enter 6 k
Calibrator outputs 660
adc vr_vc r
Read the OHM analog-to-digital conversion value
cal_param dmm_ohm 2
Write the calibration parameter 2, use read value
660 ADC_READ
ADC_READ in the previous step
Calibrator outputs 6 k
adc vr_vc r
Read the OHM analog-to-digital conversion value
cal_param dmm_ohm 3
Write the calibration parameter 3, use read value
6000 ADC_READ
ADC_READ in the previous step
drs 3
Enter 60 k
Calibrator outputs 6.6 k
adc vr_vc r
Read the OHM analog-to-digital conversion value
cal_param dmm_ohm 4
Write the calibration parameter 4, use read value
6600 ADC_READ
ADC_READ in the previous step
Calibrator outputs 60 k
adc vr_vc r
Read the OHM analog-to-digital conversion value
cal_param dmm_ohm 5
Write the calibration parameter 5, use read value
60000 ADC_READ
ADC_READ in the previous step
cal stop_dmm_ohm
Stop the OHM calibration
cal_param save
Save the calibration parameters
9.
To finish the calibration, shut down the unit under test (UUT) and remove all connections.
Table 3 is a list of the calibration points and the reference reading values used in the calibration procedure.
If any value significantly deviates from the reference range, contact the Fluke Service Center.
Ω
range
Ω
Ω
Ω
range
Ω
Ω
Ω
range
Ω
Ω
Insulation Tester
Calibration Procedure
Response
"cal start_dmm_ohm"
"drs"
ADC_READ "dmm_ohm_adc"
"dmm_ohm param"
ADC_READ "dmm_ohm_adc"
"dmm_ohm param"
"drs"
ADC_READ "dmm_ohm_adc"
"dmm_ohm param"
ADC_READ "dmm_ohm_adc"
"dmm_ohm param"
"drs"
ADC_READ "dmm_ohm_adc"
"dmm_ohm param"
ADC_READ "dmm_ohm_adc"
"dmm_ohm param"
"cal stop_dmm_ohm"
"params saved"
17