HP LTO Ultrium Technical Reference Manual page 19

Table of Contents

Advertisement

3.
The "Max temperature since cartridge loaded" value is ASCII encoded in bytes 22 29 of the
incoming data.
Example SCSI trace:
SendDiagnostic
Data Out |93 00 00 04 00 00 20 2A |
ReceiveDiagResults
Data In |93 00 00 98 30 30 30 30 31 41 31 32 20 30 30 30 |
Data In |30 31 41 31 32 20 30 30 30 30 31 41 37 34 20 30 |
Data In |30 30 30 31 39 31 37 20 0D 0A 30 30 30 30 30 30 |
Data In |30 30 20 30 30 30 30 30 30 30 30 20 30 30 30 30 |
Data In |33 32 30 30 20 30 30 30 30 30 35 30 30 20 0D 0A |
Data In |46 46 46 46 38 30 30 30 20 46 46 46 46 38 30 30 |
Data In |30 20 46 46 46 46 38 30 30 30 20 46 46 46 46 38 |
Data In |30 30 30 20 0D 0A 46 46 46 46 38 30 30 30 20 46 |
Data In |46 46 46 38 30 30 30 20 30 30 30 30 30 35 30 30 |
Data In |20 30 30 30 30 30 35 30 30 20 0D 0A |
4.
Translate each character received from ASCI to hex numbers.
For the example above: 30 30 30 30 31 41 37 34 = 00001A74h
5.
Convert this number to decimal = 6772
6.
Divide this number by 256 to give the temperature in °C = 26.4°C
|1D 11 00 00 08 00 |00008|00|00|
|1C 01 93 01 00 00 |0009C|00|00|
Volume 1: hardware integration
19