Measurement Synchronization Modes - Konica Minolta CA-410 Instruction Manual

Display color analyzer, probe + data processor
Hide thumbs Also See for CA-410:
Table of Contents

Advertisement

2. Measurement Synchronization Modes

Measurement synchronization modes are designed to select a measurement time (integration time) according to the
vertical scanning frequency of a display. A probe has the following six types of measurement synchronization modes. Select
a mode suitable for the type of the display to be measured.
Measurement
synchronization
Description of mode
mode
NTSC
Mode for measuring NTSC displays
Mode for measuring PAL and SECAM
PAL
displays
Mode for measuring while synchronizing
with the display's vertical
EXT
synchronization signal input (refer to
page 22 for how to input vertical
synchronization signals)
Mode used when the display's vertical
scanning frequency is unknown or
UNIV
cannot be input; a mode that can
measure any display
When the display's vertical
synchronization signal is known, set that
INT
information so that it can be used for
measurement.
Mode that allows you to specify the
MANUAL
integration time
On a flickering LCD, the screen is fluctuating at half the frequency of the vertical scanning frequency. When you
Memo
measure a display in this state, the luminance and the chromaticity vary if you set the same integration time as the
vertical scanning frequency.
When measuring with the measurement synchronization mode set to MANUAL, set the integration time double (or
another even multiple of) the vertical scanning frequency to ensure stable LCD measurement.
Measurement time
Vertical scanning
(Integration time)
33.3 ms
59.94 Hz
40.0 ms
50 Hz
0.50 to
(1 vertical scanning
240.00 Hz
period)
(For FMA:
×2
0.50 to 130.00 Hz)
100 ms
Settable range
(1 vertical scanning
0.50 to
period)
240.00 Hz
×2
(For FMA:
0.50 to 130.00 Hz)
Settable range
4.0 to 4000.0 ms
(For FMA:
7.7 to 4000.0 ms)
25
Display's vertical
synchronization
frequency
signal input
Not required
Not required
Required
Not required
Not required
Not required

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ca-dp40

Table of Contents