Technical specifications

10.3 OXYMAT 7 technical specifications

10.3
OXYMAT 7 technical specifications
OXYMAT 7
The information on the smallest measuring span is available on the module nameplate.
Table 10-8
General
Information on the smallest span
Power consumption
Weight of analyzer section
Electrical safety
Table 10-9
Measuring ranges
Number of measuring ranges
Parameters can be as‐
signed in the measuring
ranges
Table 10-10 Gas inlet conditions
Gas inlet conditions
Sample gas pressure
Reference gas pressure
Pressure loss between sample gas inlet and sam‐
ple gas outlet
Permissible sample gas flow range
Sample gas temperature and humidity
Optimum measuring results can only be achieved at 2000 hPa ±150 hPa above sample gas pressure.
1)
Calibrations are required for any larger deviations.
90
General technical specifications
Warm-up phase
Operating phase
Measuring ranges
Smallest possible spans 0.5 %, 1%, 2 % or 5 % O
Largest possible span
for non-combustible
gases
for combustible gases
up to gas mixtures
which are occasionally
explosive
See module rating plate or device certificate
≤ 95 W
≤ 35 W
Approx. 5.5 kg (standard version)
In accordance with DIN EN 61010-1
Max. 4; parameters can be assigned freely
2
100% O
2
500 to 2500 hPa (abs.)
800 ... 1100 hPa (abs.)
2000 .. 4000 hPa above the sample gas pressure
Max. permissible: 5000 hPa (abs.)
< 100 hPa at 1 l/min
18 to 60 l/h (0.3 to 1 l/min)
0 to 60 °C with rel. humidity < 90 %, condensation
in the gas path must be prevented (it must be taken
into account that moisture content may cause
measurement errors)
Operating Instructions, 05/2015, A5E35640463-02