Wiring
6.5 Field control unit
3. Guide the connecting cables through the corresponding cable glands.
– Make sure in the enclosure that the cables are routed tidily and force-free.
– Use the clips provided in the enclosure.
4. Connect the power cable to the envisaged terminals of the terminal block.
5. Connect the field module cable according to the terminal assignments.
6. Close the cable gland. Tightening torque: 0.8 Nm
7. Close all open covers of the field control unit. Tightening torque: 3.5 Nm
6.5.4

Detail of Ethernet connection

Ethernet connection
● Standard Ethernet interface (bottom): Ethernet RJ 45, 100 Mbit
An overview of the terminal assignments can be found on the inside of the cover when removed.
Figure 6-12
Table 6-2
Connection of Ethernet cable shielding
The Ethernet cable shielding is connected on the customer side underneath the terminal.
Connect the Ethernet cable as shown in the following example.
62
Connection on device side
Connection on customer side
Ethernet connection
Ethernet cable connection
Color
White
Blue
Yellow
Orange
Terminal no.
3 (Rx+)
6 (Rx-)
4 (Tx+)
5 (Tx-)
Operating Instructions, 05/2015, A5E35640463-02
Info
Length of lay < 6 mm
Field device