Ex py safety equipment

A.1.1

Overview

Example of Ex p safety equipment
A
Ex interface relay SR852
B
Pressure regulating valve for purging gas with gauge
C
Proportional valve
Purging gas connection 2.5 ... 5 bar (rel.)
Connection to Ex p enclosure (SIPROCESS GA700); conductor cross-section 8 mm, max. length 10 m
Connection from Ex p enclosure (SIPROCESS GA700); conductor cross-section 8 mm, max. length 10 m
Connection of power cable; fusing 16 A
Exhaust air outlet. If appropriate connection of a 1" line, max. length 20 m
Figure A-1
Devices in explosion-proof design
Compact Operating Instructions, 05/2018, A5E35134047-03
Ex p safety equipment, Gönnheimer Elektronik GmbH
A
D
Ex p controller F870S
E
Ex power relay SR853
F
Ground connection