8.8 Storage and transportation

8.8.1 Prepare the instrument for storage

46
Option
Live measure
Reagent
replacement
Maintenance
schedule
Maintenance
summary
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
1. Remove the sensor from the sensor holder and disconnect it from the sensor port on
the rear of the instrument.
2. Clean the sensor. Refer to
3. Put the sensor in a storage tube that contains a storage solution. Refer to the sensor
documentation for storage instructions.
4. From the home screen, select Maintenance > Syringe activation and empty the
contents of the syringe.
5. If a second syringe is installed, push Toggle to select it and empty the contents.
6. Set the instrument power to off.
7. Carefully remove the tubes from the titrant bottle. If pumps are installed, remove the
tubes from the reagent bottles. Seal the bottles with the original bottle cap.
8. Flush the tubes with water or an applicable cleaning solution (refer to the application
note for more information).
9. Drain the used beakers and clean with deionized water.
10. If storing the instrument for more than a week, remove the syringe from the
instrument (refer to
applicable cleaning solution and store in its box.
Description
This option is not available for all sensors. The option shows continuous
measurement data with connected sensors to quickly check
measurements. The installed applications and the automatic additions to
the sample are set to off. Continuous measurements are not
temperature compensated, so measurement differences may occur in
the same sample between continuous measurements and
measurements that use installed applications with temperature
compensation.
Use this option to replace the reagents. Obey the instructions on the
display.
See the list of maintenance tasks. Push Edit to change the default
value.
See the number of days remaining for maintenance tasks. After doing a
task, push Reset to set the number of days remaining to the default
value.
C A U T I O N
Clean the sensor
Install the syringe
on page 12). Flush the syringe with water or an
on page 41.