Section 3 Installation; Installation Guidelines; Connect To Ac Power - Hach TitraLab AT1102 User Manual

Hide thumbs Also See for TitraLab AT1102:
Table of Contents

Advertisement

Section 3

3.1 Installation guidelines

3.2 Connect to AC power

Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
This is a class A product. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated disturbances. In a
domestic environment this product may cause radio interference in which case the user may be
required to take adequate measures.
Network and access point security is the responsibility of the customer that uses the wireless
instrument. The manufacturer will not be liable for any damages, inclusive however not limited to
indirect, special, consequential or incidental damages, that have been caused by a gap in, or
breach of network security.
The instrument is available in different configurations (refer to
manual supplies instructions for the installation of an instrument with one syringe and one
pump. Adjust the installation procedure as applicable to accommodate the number of
syringes and pumps in the instrument.
This instrument is for indoor use only.
The power supply connector on the rear panel must be easily accessible so the
power can be disconnected quickly in case of emergency.
Keep the instrument away from temperature extremes, including heaters, direct
sunlight and other heat sources.
Put the instrument on a stable and level surface in a well ventilated place.
Make sure that there is at least 15 cm (6 in.) of space on all sides of the instrument to
prevent electrical parts from overheating.
Do not operate or keep the instrument in dusty, damp or wet locations.
Always keep the surface of the instrument and all accessories dry and clean.
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a
Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the
equipment to its main power source.
Electrical shock and fire hazards. Make sure that the supplied cord and non‐locking plug
meet the applicable country code requirements.
Fire hazard. Use only the external power supply that is specified for this instrument.
C A U T I O N
N O T I C E
N O T I C E
D A N G E R
C A U T I O N
W A R N I N G
Table 2
on page 8). This
11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents