About this document Proline Cubemass 300 Modbus RS485 About this document Document function The document is part of the Operating Instructions and serves as a reference for parameters, providing a detailed explanation of each individual parameter of the Expert operating menu.
Page 5
Proline Cubemass 300 Modbus RS485 About this document Operating menu for operators and maintenances Language Language Operatation Parameter 1 Parameter n Submenu 1 Submenu n Setup Device tag Wizard 1 / Parameter 1 Wizard n / Parameter n Advanced setup...
About this document Proline Cubemass 300 Modbus RS485 1.3.2 Structure of a parameter description The individual parts of a parameter description are described in the following section: Complete parameter name Write-protected parameter = Navigation Navigation path to the parameter via the local display (direct access code) or web browser...
Proline Cubemass 300 Modbus RS485 About this document 1.4.2 Symbols in graphics Symbol Meaning Symbol Meaning 1, 2, 3 ... Item numbers A, B, C, ... Views A-A, B-B, C-C, ... Sections Documentation 1.5.1 Standard documentation Operating Instructions Measuring device...
Overview of the Expert operating menu Proline Cubemass 300 Modbus RS485 Overview of the Expert operating menu The following table provides an overview of the menu structure of the expert operating menu and its parameters. The page reference indicates where the associated description of the submenu or parameter can be found.
Page 9
Proline Cubemass 300 Modbus RS485 Overview of the Expert operating menu I/O 1 type (3901–1) → 102 Apply I/O config (3907) → 102 Alteration code (2762) → 103 ‣ Input → 103 ‣ Current input 1 to n →...
Page 10
Overview of the Expert operating menu Proline Cubemass 300 Modbus RS485 Operating time (0652) → 185 ‣ Diagnostic list → 186 ‣ Event logbook → 190 ‣ Device info → 192 ‣ → 196 Mainboard module ‣...
Proline Cubemass 300 Modbus RS485 Description of device parameters Description of device parameters In the following section, the parameters are listed according to the menu structure of the local display. Specific parameters for the operating tools are included at the appropriate points in the menu structure.
Page 12
Description of device parameters Proline Cubemass 300 Modbus RS485 00914-2 A0029414 Direct access code Note the following when entering the direct access code: • The leading zeros in the direct access code do not have to be entered. Example: Enter "914" instead of "00914"...
Proline Cubemass 300 Modbus RS485 Description of device parameters Access status Navigation Expert → Access status (0005) Description Displays the access authorization to the parameters via the local display, Web browser or operating tool. User interface • Operator • Maintenance...
Description of device parameters Proline Cubemass 300 Modbus RS485 3.1.1 "Display" submenu Navigation Expert → System → Display ‣ Display Display language (0104) → 15 Format display (0098) → 15 → 18 Value 1 display (0107) 0% bargraph 1 (0123) →...
Page 15
Proline Cubemass 300 Modbus RS485 Description of device parameters Display language Navigation Expert → System → Display → Display language (0104) Prerequisite A local display is provided. Description Use this function to select the configured language on the local display.
Page 16
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description The display format (size, bar graph etc.) and number of measured values displayed simultaneously (1 to 4) can be configured. This setting only applies to normal operation. • The Value 1 display parameter (→ 18) to Value 4 display parameter (→...
Page 17
Proline Cubemass 300 Modbus RS485 Description of device parameters Possible measured values shown on the local display: "1 value, max." option X X X X X X X 900.00 kg/h A0013099 "Bargr. + 1 value" option X X X X X X X 900.00 kg/h...
Page 18
Description of device parameters Proline Cubemass 300 Modbus RS485 Value 1 display Navigation Expert → System → Display → Value 1 display (0107) Prerequisite A local display is provided. Description Use this function to select one of the measured values to be shown on the local display.
Page 19
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description If several measured values are displayed at once, the measured value selected here will be the first value to be displayed. The value is only displayed during normal operation.
Page 20
Description of device parameters Proline Cubemass 300 Modbus RS485 100% bargraph 1 Navigation Expert → System → Display → 100% bargraph 1 (0125) Prerequisite A local display is provided. Description Use this function to enter the 100% bar graph value to be shown on the display for the measured value 1.
Page 21
Proline Cubemass 300 Modbus RS485 Description of device parameters Description Use this function to select one of the measured values to be shown on the local display. Selection • None • Mass flow • Volume flow • Correct.vol.flow • Target mass flow •...
Page 22
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description If several measured values are displayed at once, the measured value selected here will be the second value to be displayed. The value is only displayed during normal operation.
Page 23
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description If several measured values are displayed at once, the measured value selected here will be the third value to be displayed. The value is only displayed during normal operation.
Page 24
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description The Format display parameter (→ 15) is used to specify that the measured value is to be displayed as a bar graph. User entry The unit of the displayed measured value is taken from the System units submenu (→...
Page 25
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description If several measured values are displayed at once, the measured value selected here will be the fourth value to be displayed. The value is only displayed during normal operation.
Page 26
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description This type of alternating display only occurs automatically if the number of measured values defined exceeds the number of values the selected display format can display simultaneously. • The Value 1 display parameter (→ 18) to Value 4 display parameter (→ 24) are used to specify which measured values are shown on the local display.
Page 27
Proline Cubemass 300 Modbus RS485 Description of device parameters XXXXXXXXX A0029422 Position of the header text on the display Selection • Device tag Is defined in the Device tag parameter (→ 193). • Free text Is defined in the Header text parameter (→ 27).
Description of device parameters Proline Cubemass 300 Modbus RS485 Description Use this function to select the decimal separator. Selection • . (point) • , (comma) Factory setting . (point) Contrast display Navigation Expert → System → Display → Contrast display (0105) Prerequisite A local display is provided.
Page 29
Proline Cubemass 300 Modbus RS485 Description of device parameters Last backup → 29 Config. managem. → 29 Backup state → 30 Compar. result → 30 Operating time Navigation Expert → System → Config. backup → Operating time (0652) Description Use this function to display the length of time the device has been in operation.
Page 30
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Selection Options Description Cancel No action is executed and the user exits the parameter. Execute backup A backup copy of the current device configuration is saved from the HistoROM backup to the memory of the device.
Proline Cubemass 300 Modbus RS485 Description of device parameters • Backup corrupt • Check not done • Dataset incomp. Factory setting Check not done Additional information Description The comparison is started via the Compare option in the Config. managem. parameter (→ 29).
Page 32
Description of device parameters Proline Cubemass 300 Modbus RS485 Alarm delay Navigation Expert → System → Diagn. handling → Alarm delay (0651) Description Use this function to enter the time interval until the device generates a diagnostic message.
Page 34
Description of device parameters Proline Cubemass 300 Modbus RS485 Diagnostic no. 046 (Sensor limit) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 046 (0709) Description Option for changing the diagnostic behavior of the diagnostic message 046 Sensor limit.
Page 35
Proline Cubemass 300 Modbus RS485 Description of device parameters Diagnostic no. 374 (Sensor electron.) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 374 (0710) Description Option for changing the diagnostic behavior of the diagnostic message 374 Sensor electron..
Page 36
Description of device parameters Proline Cubemass 300 Modbus RS485 Diagnostic no. 442 (Freq. output 1 to n) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 442 (0658) Prerequisite The measuring device has a pulse/frequency/switch output.
Page 37
Proline Cubemass 300 Modbus RS485 Description of device parameters Selection • Off • Alarm • Warning • Logbook only Factory setting Warning Additional information For a detailed description of the options available, see → 32 Diagnostic no. 543 (Double pulse out) ...
Page 38
Description of device parameters Proline Cubemass 300 Modbus RS485 Diagnostic no. 831 (Sensor temp.) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 831 (0641) Description Use this function to change the diagnostic behavior of the diagnostic message 831 Sensor temp..
Page 39
Proline Cubemass 300 Modbus RS485 Description of device parameters Diagnostic no. 834 (Process temp.) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 834 (0700) Description Use this function to change the diagnostic behavior of the diagnostic message 834 Process temp..
Page 40
Description of device parameters Proline Cubemass 300 Modbus RS485 Diagnostic no. 912 (Medium inhomog.) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 912 (0703) Description Option for changing the diagnostic behavior of the diagnostic message 912 Medium inhomog..
Proline Cubemass 300 Modbus RS485 Description of device parameters Diagnostic no. 948 (Oscill. damping) Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 948 (0744) Description Option for changing the diagnostic behavior of the diagnostic message 948 Oscill.
Page 42
Once the access code has been defined, write-protected parameters can only be modified if the access code is entered in the Ent. access code parameter (→ 13). If you lose the access code, please contact your Endress+Hauser sales organization. User entry A message is displayed if the access code is not in the input range.
Page 43
Character string comprising numbers, letters and special characters Factory setting 0x00 Additional information Description For a reset code, contact your Endress+Hauser service organization. User entry The reset code can only be entered via: • Web browser • DeviceCare, FieldCare (via interface CDI RJ45) •...
Page 44
Additional information Description If a measuring device was ordered with an additional software option, the activation code is programmed in the device at the factory. User entry To activate a software option subsequently, please contact your Endress+Hauser sales organization. NOTE! Endress+Hauser...
Page 45
▸ Before you enter a new activation code, make a note of the current activation code . ▸ Enter the new activation code provided by Endress+Hauser when the new software option was ordered. ▸ Once the activation code has been entered, check if the new software option is displayed in the SW option overv.
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description Displays all the options that are available if ordered by the customer. "Extend. HistoROM" option Order code for "Application package", option EA "Extended HistoROM" "HBT Verification" option and "HBT Monitoring" option Order code for "Application package", option EB "Heartbeat Verification + Monitoring"...
Page 47
Proline Cubemass 300 Modbus RS485 Description of device parameters ‣ Input values → 54 ‣ Output values → 55 "Process variab." submenu Navigation Expert → Sensor → Measured val. → Process variab. ‣ Process variab. Mass flow (1838) →...
Page 48
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Dependency The unit is taken from the Mass flow unit parameter (→ 60) Volume flow Navigation Expert → Sensor → Measured val. → Process variab. → Volume flow (1847) Description Displays the volume flow currently calculated.
Page 49
Proline Cubemass 300 Modbus RS485 Description of device parameters Ref.density Navigation Expert → Sensor → Measured val. → Process variab. → Ref.density (1852) Description Displays the reference density currently calculated. User interface Signed floating-point number Additional information Dependency The unit is taken from the Ref. dens. unit parameter (→ 66)
Page 50
Description of device parameters Proline Cubemass 300 Modbus RS485 Description Displays the concentration currently calculated. User interface Signed floating-point number Additional information Dependency The unit is taken from the Concentr. unit parameter (0613). Target mass flow Navigation Expert → Sensor → Measured val. → Process variab. → Target mass flow (1864)
Page 51
Proline Cubemass 300 Modbus RS485 Description of device parameters Targ.corr.vol.fl Navigation Expert → Sensor → Measured val. → Process variab. → Targ.corr.vol.fl (1893) Prerequisite With the following conditions: • Order code for "Application package", option ED "Concentration" • The WT-% option or the User conc. option is selected in the Concentr. unit parameter.
Page 52
Description of device parameters Proline Cubemass 300 Modbus RS485 User interface Signed floating-point number Additional information Dependency The unit is taken from the Volume flow unit parameter (→ 61) Carrier vol. fl. Navigation Expert → Sensor → Measured val. → Process variab. → Carrier vol. fl. (1896)
Page 53
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description As it is only possible to display a maximum of 7 digits in the operating tool, the current counter value is the sum of the totalizer value and the overflow value from the Tot.
Page 54
Description of device parameters Proline Cubemass 300 Modbus RS485 "Input values" submenu Navigation Expert → Sensor → Measured val. → Input values ‣ Input values ‣ Current input 1 to n → 54 ‣ Val.stat.inp. 1 to n →...
Page 55
Proline Cubemass 300 Modbus RS485 Description of device parameters "Value status input 1 to n" submenu Navigation Expert → Sensor → Measured val. → Input values → Val.stat.inp. 1 to n ‣ Val.stat.inp. 1 to n Val.stat.inp. (1353–1 to n) →...
Page 56
Description of device parameters Proline Cubemass 300 Modbus RS485 Output curr. 1 to n Navigation Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n → Output curr. 1 to n (0361–1 to n) Description Displays the current value currently calculated for the current output.
Page 57
Proline Cubemass 300 Modbus RS485 Description of device parameters Pulse output 1 to n Navigation Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Pulse output 1 to n (0456–1 to n) Prerequisite In the Operating mode parameter (→...
Page 58
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information User interface • Open The switch output is not conductive. • Closed The switch output is conductive. "Relay output 1 to n" submenu Navigation Expert → Sensor → Measured val. → Output values → Relay output 1 to n ‣...
Proline Cubemass 300 Modbus RS485 Description of device parameters Max. cycles no. Navigation Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Max. cycles no. (0817–1 to n) Description Displays the maximum number of guaranteed switch cycles.
Page 60
Description of device parameters Proline Cubemass 300 Modbus RS485 Volume unit (0563) → 63 Cor.volflow unit (0558) → 64 Corr. vol. unit (0575) → 64 Density unit (0555) → 65 → 66 Ref. dens. unit (0556) Temperature unit (0557) →...
Page 61
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Result The selected unit applies for: • Target mass flow parameter (→ 50) • Carrier mass fl. parameter (→ 50) • Mass flow parameter (→ 47) Selection For an explanation of the abbreviated units: →...
Page 62
Description of device parameters Proline Cubemass 300 Modbus RS485 Selection SI units US units Imperial units • cm³/s • af/s • gal/s (imp) • cm³/min • af/min • gal/min (imp) • cm³/h • af/h • gal/h (imp) • cm³/d • af/d •...
Page 63
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Result The selected unit applies for: Volume flow parameter (→ 48) Selection For an explanation of the abbreviated units: → 232 Customer-specific units The unit for the customer-specific volume is specified in the Volume text parameter (→...
Page 64
Description of device parameters Proline Cubemass 300 Modbus RS485 Cor.volflow unit Navigation Expert → Sensor → System units → Cor.volflow unit (0558) Description Use this function to select the unit for the corrected volume flow. Selection SI units...
Page 65
Proline Cubemass 300 Modbus RS485 Description of device parameters Density unit Navigation Expert → Sensor → System units → Density unit (0555) Description Use this function to select the unit for the density. Selection SI units US units Imperial units •...
Page 66
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Result The selected unit applies for: • Density setpt 1 parameter • Density setpt 2 parameter • Density parameter (→ 48) Selection • SD = specific density The specific density is the ratio of the medium density to the water density at a water temperature of +4 °C (+39 °F), +15 °C (+59 °F), +20 °C (+68 °F).
Page 67
Proline Cubemass 300 Modbus RS485 Description of device parameters Temperature unit Navigation Expert → Sensor → System units → Temperature unit (0557) Description Use this function to select the unit for the temperature. Selection SI units US units •...
Page 68
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Result The unit is taken from: • Pressure value parameter (→ 88) • External press. parameter (→ 88) • Pressure value parameter (→ 49) Selection For an explanation of the abbreviated units: → 232 Date/time format ...
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Result The defined unit is shown as an option in the choose list of the following parameters: • Volume flow unit parameter (→ 61) • Volume unit parameter (→ 63) Example If the text GLAS is entered, the choose list of the Volume flow unit parameter (→...
Proline Cubemass 300 Modbus RS485 Description of device parameters User entry Max. 10 characters such as letters, numbers or special characters (@, %, /) Factory setting User mass Additional information Result The defined unit is shown as an option in the choose list of the following parameters: •...
Description of device parameters Proline Cubemass 300 Modbus RS485 Corr. vol. text Navigation Expert → Sensor → System units → User-spec. units → Corr. vol. text (0592) Description Use this function to enter a text for the user-specific unit of the corrected volume and corrected volume flow.
Page 73
Proline Cubemass 300 Modbus RS485 Description of device parameters Density text Navigation Expert → Sensor → System units → User-spec. units → Density text (0570) Description Use this function to enter a text or the user-specific unit of density.
Page 74
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting User enth. Additional information Result Example If the text CAL is entered, the choose list of the Cal. value unit parameter shows the following options: • CAL/Nm3 • CAL/m3 •...
Page 75
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Result The defined unit is shown as an option in the choose list of the following parameters: • Energy unit parameter • Energy flow unit parameter Example If the text W is entered, the choose list of the Energy flow unit parameter shows the following options: •...
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Result The defined unit is shown as an option in the choose list of the Pressure unit parameter (→ 67). Pressure offset Navigation Expert → Sensor → System units → User-spec. units → Pressure offset (0580) Description Use this function to enter the offset for adapting the user-specific pressure unit.
Page 77
Proline Cubemass 300 Modbus RS485 Description of device parameters ‣ Low flow cut off → 79 ‣ Partial pipe det → 82 Flow damping Navigation Expert → Sensor → Process param. → Flow damping (1802) Description Use this function to enter a time constant for flow damping (PT1 element).
Page 78
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description The damping is performed by a PT1 element User entry • Value = 0: no damping • Value > 0: damping is increased Damping is switched off if 0 is entered (factory setting).
Page 79
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description Flow override is active • The diagnostic message diagnostic message C453 Flow override is displayed. • Output values – Temperature: proceeding output – Totalizers 1-3: Stop being totalized Positive zero return can also be enabled via the Status input: Assign stat.inp.
Page 80
Description of device parameters Proline Cubemass 300 Modbus RS485 Description Use this function to enter a switch-on value for low flow cut off. Low flow cut off is activated if the value entered is not equal to 0 → 80.
Page 81
Proline Cubemass 300 Modbus RS485 Description of device parameters Pres. shock sup. Navigation Expert → Sensor → Process param. → Low flow cut off → Pres. shock sup. (1806) Prerequisite One of the following options is selected in the Assign variable parameter (→ 79): •...
Page 82
Description of device parameters Proline Cubemass 300 Modbus RS485 A0012888 Flow Time Drip Pressure shock Pressure shock suppression active as specified by the time entered Pressure shock suppression inactive Valve closes Flow falls below the on-value of the low flow cut off: pressure shock suppression is activated...
Page 83
Proline Cubemass 300 Modbus RS485 Description of device parameters Assign variable Navigation Expert → Sensor → Process param. → Partial pipe det → Assign variable (1860) Description Use this function to select a process variable to detect empty or partially filled measuring tubes.
Page 84
Description of device parameters Proline Cubemass 300 Modbus RS485 Description Use this function to enter an upper limit value to enable detection of empty or partially filled measuring tubes. If the measured density exceeds this value, detection is enabled. User entry...
Proline Cubemass 300 Modbus RS485 Description of device parameters diagnostic message S862 Partly filled. In the case of non-homogeneous media or air pockets, the damping of the measuring tubes increases. User entry • Damping is disabled if 0 is entered (factory setting).
Page 86
Description of device parameters Proline Cubemass 300 Modbus RS485 Selection • Liquid • Gas Factory setting Liquid Select gas type Navigation Expert → Sensor → Measurement mode → Select gas type (6074) Prerequisite The Gas option is selected in the Select medium parameter (→ 85).
Proline Cubemass 300 Modbus RS485 Description of device parameters User entry 1 to 99 999.9999 m/s Factory setting 415.0 m/s Temp. coeff. SV Navigation Expert → Sensor → Measurement mode → Temp. coeff. SV (6181) Prerequisite The Others option is selected in the Select gas type parameter (→ 86).
Page 88
Description of device parameters Proline Cubemass 300 Modbus RS485 Selection • Off • Fixed value • External value • Current input 1 • Current input 2 Factory setting Additional information Selection • Fixed value A fixed pressure value is used for compensation: Pressure value parameter (→ 88) •...
Page 89
Proline Cubemass 300 Modbus RS485 Description of device parameters Temp.corr.source Navigation Expert → Sensor → External comp. → Temp.corr.source (6184) Description Use this function to select the temperature mode. Selection • Internal value • External value • Current input 1 •...
Description of device parameters Proline Cubemass 300 Modbus RS485 Selection • Internal value • External value Factory setting Internal value 3.2.6 "Calculated value" submenu Navigation Expert → Sensor → Calculated value ‣ Calculated value ‣ → 90 Corr. vol.flow.
Page 91
Proline Cubemass 300 Modbus RS485 Description of device parameters Factory setting Calc ref density Additional information Selection The Ref. dens API 53 option is suitable only for applications involving LPG , where the flow rate is measured on the basis of the corrected volume flow.
Page 92
Description of device parameters Proline Cubemass 300 Modbus RS485 Ref. temperature Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Ref. temperature (1816) Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→ 90).
Proline Cubemass 300 Modbus RS485 Description of device parameters Square exp coeff Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Square exp coeff (1818) Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→ 90) parameter.
Page 94
Description of device parameters Proline Cubemass 300 Modbus RS485 "Zero point adj." submenu • It is generally not necessary to perform zero point adjustment. • However, this function may be needed in some applications with low flow and strict accuracy requirements.
Page 95
Proline Cubemass 300 Modbus RS485 Description of device parameters Progress Navigation Expert → Sensor → Sensor adjustm. → Zero point adj. → Progress (2808) Description The progress of the process is indicated. User interface 0 to 100 % "Variable adjust" submenu Navigation ...
Page 96
Description of device parameters Proline Cubemass 300 Modbus RS485 User entry Signed floating-point number Factory setting 0 kg/s Additional information Description Corrected value = (factor × value) + offset Mass flow factor Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow factor (1832) Description Use this function to enter a quantity factor (without time) for the mass flow.
Page 97
Proline Cubemass 300 Modbus RS485 Description of device parameters Factory setting Additional information Description Corrected value = (factor × value) + offset Density offset Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Density offset (1848) Description Use this function to enter the zero point shift for the density trim.
Page 98
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description Corrected value = (factor × value) + offset Corr. vol factor Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol factor (1867) Description Use this function to enter a quantity factor (without time) for the corrected volume flow.
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description Corrected value = (factor × value) + offset Temp. offset Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. offset (1870) Description Use this function to enter the zero point shift for the temperature trim. The temperature unit on which the shift is based is K.
Page 100
Description of device parameters Proline Cubemass 300 Modbus RS485 Nominal diameter (2807) → 100 C0 to 5 (6022) → 101 Cal. factor Navigation Expert → Sensor → Calibration → Cal. factor (6025) Description Displays the current calibration factor for the sensor.
Proline Cubemass 300 Modbus RS485 Description of device parameters C0 to 5 Navigation Expert → Sensor → Calibration → C0 to 5 (6022) Description Displays the current density coefficients C0 to 5 of the sensor. User interface Signed floating-point number Factory setting "I/O configuration"...
Page 102
Description of device parameters Proline Cubemass 300 Modbus RS485 User interface • Not plugged • Invalid • Not configurable • Configurable • MODBUS Additional information "Not plugged" option The I/O module is not plugged in. "Invalid" option The I/O module is not plugged correctly.
Proline Cubemass 300 Modbus RS485 Description of device parameters Selection • No • Yes Factory setting Alteration code Navigation Expert → I/O config. → Alteration code (2762) Description Use this function to enter the ordered activation code to activate the I/O configuration change.
Page 104
Description of device parameters Proline Cubemass 300 Modbus RS485 Failure mode (1601–1 to n) → 106 Failure value (1602–1 to n) → 106 Terminal no. Navigation Expert → Input → Current input 1 to n → Terminal no. (1611–1 to n) Description Displays the terminal numbers used by the current input module.
Page 105
Proline Cubemass 300 Modbus RS485 Description of device parameters Factory setting Country-specific: • 4...20 mA NAMUR • 4...20 mA US Additional information Examples Sample values for the current range: Current span parameter (→ 111) 0/4 mA value Navigation ...
Description of device parameters Proline Cubemass 300 Modbus RS485 Failure mode Navigation Expert → Input → Current input 1 to n → Failure mode (1601–1 to n) Description Use this function to select the input behavior when measuring a current outside the configured Current span parameter (→...
Page 107
Proline Cubemass 300 Modbus RS485 Description of device parameters Active level (1351–1 to n) → 108 Response time (1354–1 to n) → 108 Terminal no. Navigation Expert → Input → Status input 1 to n → Terminal no. (1358–1 to n) Description Displays the terminal numbers used by the status input module.
Description of device parameters Proline Cubemass 300 Modbus RS485 Val.stat.inp. Navigation Expert → Input → Status input 1 to n → Val.stat.inp. (1353–1 to n) Description Displays the current input signal level. User interface • High • Low Active level ...
Proline Cubemass 300 Modbus RS485 Description of device parameters ‣ Relay output 1 to n → 148 ‣ Double pulse out → 155 3.5.1 "Current output 1 to n" submenu Navigation Expert → Output → Curr.output 1 to n ‣...
Page 110
Description of device parameters Proline Cubemass 300 Modbus RS485 User interface • Not used • 24-25 (I/O 2) • 22-23 (I/O 3) Additional information "Not used" option The current output module does not use any terminal numbers. Signal mode ...
Page 111
Proline Cubemass 300 Modbus RS485 Description of device parameters • Water mass flow • Oil volume flow • Water vol. flow • Oil corr.vol.fl. • Water corr.v.fl. • Concentration • Temperature • Carr. pipe temp. • Electronic temp. • Osc. freq. 0 •...
Page 112
Description of device parameters Proline Cubemass 300 Modbus RS485 I [mA] A0034351 Current span for process value Lower level for signal on alarm Upper level for signal on alarm Selection Options 4...20 mA NAMUR 3.8 to 20.5 mA < 3.6 mA >...
Page 113
Proline Cubemass 300 Modbus RS485 Description of device parameters Factory setting Country-specific: • 0 kg/h • 0 lb/min Additional information Description Positive and negative values are permitted depending on the process variable assigned in the Assign curr. parameter (→ 110). In addition, the value can be greater than or smaller than the value assigned for the 20 mA current in the 20 mA value parameter (→...
Page 114
Description of device parameters Proline Cubemass 300 Modbus RS485 I [mA] A0013758 Current Flow Value assigned to the 0/4 mA current Forward flow Reverse flow The current output signal is independent of the direction of flow (absolute amount of the measured variable).
Page 115
Proline Cubemass 300 Modbus RS485 Description of device parameters smaller than the value assigned for the 0/4 mA current in the 0/4 mA value parameter (→ 112). Dependency The unit depends on the process variable selected in the Assign curr. parameter (→...
Page 116
Description of device parameters Proline Cubemass 300 Modbus RS485 Description Use this function to select the measuring mode for the current output. Selection • Forward flow • Forward/Reverse • Rev. flow comp. Factory setting Forward flow Additional information Description The process variable that is assigned to the current output via the Assign curr.
Page 117
Proline Cubemass 300 Modbus RS485 Description of device parameters Flow values can aggregate in the buffer in the event of prolonged and unwanted fluid reverse flow. However, these flows are not taken into consideration by the current output configuration, i.e. the reverse flow is not compensated.
Page 118
Description of device parameters Proline Cubemass 300 Modbus RS485 I [mA] A0028093 Current Time With Rev. flow comp. option Flow components outside the span are buffered, balanced and output after a maximum delay of 60 s. I [mA] S = A...
Page 119
Proline Cubemass 300 Modbus RS485 Description of device parameters A0028098 Flow Time Lower range value (value assigned to 0/4 mA current) Upper range value (value assigned to 20 mA current) With Forward flow option • a (—): The flow components outside the scaled measuring range cannot be taken into account for signal output.
Page 120
Description of device parameters Proline Cubemass 300 Modbus RS485 Damping out. 1 to n Navigation Expert → Output → Curr.output 1 to n → Damping out. 1 to n (0363–1 to n) Prerequisite One of the following options is selected in the Assign curr. parameter (→ 110): •...
Page 121
Proline Cubemass 300 Modbus RS485 Description of device parameters Response time Navigation Expert → Output → Curr.output 1 to n → Response time (0378–1 to n) Prerequisite One of the following options is selected in the Assign curr. parameter (→ 110): •...
Page 122
Description of device parameters Proline Cubemass 300 Modbus RS485 Failure mode Navigation Expert → Output → Curr.output 1 to n → Failure mode (0364–1 to n) Prerequisite One of the following options is selected in the Assign curr. parameter (→ 110): •...
Page 123
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description This setting does not affect the failsafe mode of other outputs and totalizers. This is specified in separate parameters. "Min." option The current output adopts the value of the lower level for signal on alarm.
Description of device parameters Proline Cubemass 300 Modbus RS485 Measur. curr. 1 to n Navigation Expert → Output → Curr.output 1 to n → Measur. curr. 1 to n (0366–1 to n) Description Use this function to display the actual measured value of the output current.
Page 125
Proline Cubemass 300 Modbus RS485 Description of device parameters Response time (0491–1 to n) → 138 Failure mode (0451–1 to n) → 139 Failure freq. (0474–1 to n) → 140 Output freq. 1 to n (0471–1 to n) →...
Page 126
Description of device parameters Proline Cubemass 300 Modbus RS485 Signal mode Navigation Expert → Output → PFS output 1 to n → Signal mode (0490–1 to n) Description Use this function to select the signal mode for the pulse/frequency/switch output.
Page 127
Proline Cubemass 300 Modbus RS485 Description of device parameters U[V] B < P A0026883 5 Quantity-proportional pulse (pulse value) with pulse width to be configured Pulse width entered Pauses between the individual pulses "Frequency" option Flow-proportional frequency output with 1:1 on/off ratio...
Page 128
Description of device parameters Proline Cubemass 300 Modbus RS485 U[V] A0026885 8 Alarm, low level Assign pulse 1 to n Navigation Expert → Output → PFS output 1 to n → Assign pulse 1 to n (0460–1 to n) Prerequisite The Pulse option is selected in the Operating mode parameter (→...
Page 129
Proline Cubemass 300 Modbus RS485 Description of device parameters Value per pulse Navigation Expert → Output → PFS output 1 to n → Value per pulse (0455–1 to n) Prerequisite In the Operating mode parameter (→ 126), the Pulse option is selected, and one of the following options is selected in the Assign pulse parameter (→...
Page 130
Description of device parameters Proline Cubemass 300 Modbus RS485 U[V] B < P U[V] B = P A0026882 Pulse width entered Pauses between the individual pulses Example • Pulse value: 0.1 g • Pulse width: 0.1 ms • f : 1 / (2 × 0.1 ms) = 5 kHz •...
Page 131
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Selection • Forward flow Positive flow is output, negative flow is not output. • Forward/Reverse Positive and negative flow are output (absolute value), but a distinction is not made between positive and negative flow.
Page 132
Description of device parameters Proline Cubemass 300 Modbus RS485 Actual value option is only recommended if it can be guaranteed that all possible alarm conditions will not affect the measurement quality. Pulse output 1 to n Navigation Expert → Output → PFS output 1 to n → Pulse output 1 to n (0456–1 to n) Prerequisite In the Operating mode parameter (→...
Page 135
Proline Cubemass 300 Modbus RS485 Description of device parameters Val. at min.freq Navigation Expert → Output → PFS output 1 to n → Val. at min.freq (0476–1 to n) Prerequisite In the Operating mode parameter (→ 126), the Frequency option is selected, and one of the following options is selected in the Assign freq.
Page 137
Proline Cubemass 300 Modbus RS485 Description of device parameters • Osc.damp.fluct 0 • Signal asymmetry • Exc. current 0 Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil. damping and Signal asymmetry: Value 1 display parameter (→ 18) Description Use this function to select the measuring mode for the frequency output.
Page 138
Description of device parameters Proline Cubemass 300 Modbus RS485 Description Use this function to enter a time constant for the reaction time of the output signal to fluctuations in the measured value. User entry 0 to 999.9 s Factory setting 0.0 s...
Page 139
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description The response time is made up of the time specified for the following dampings: • Damping of pulse/frequency/switch output → 120 • Depending on the measured variable assigned to the output.
Page 140
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Selection • Actual value In the event of a device alarm, the frequency output continues on the basis of the current flow measurement. The device alarm is ignored. • Defined value In the event of a device alarm, the frequency output continues on the basis of a predefined value.
Page 141
Proline Cubemass 300 Modbus RS485 Description of device parameters Output freq. 1 to n Navigation Expert → Output → PFS output 1 to n → Output freq. 1 to n (0471–1 to n) Prerequisite In the Operating mode parameter (→ 126), the Frequency option is selected.
Page 142
Description of device parameters Proline Cubemass 300 Modbus RS485 Assign diag. beh Navigation Expert → Output → PFS output 1 to n → Assign diag. beh (0482–1 to n) Prerequisite • In the Operating mode parameter (→ 126), the Switch option is selected.
Page 143
Proline Cubemass 300 Modbus RS485 Description of device parameters • GSVa • NSV flow • NSVa • S&W volume flow • Water cut • Oil density • Water density • Oil mass flow • Water mass flow • Oil volume flow •...
Page 144
Description of device parameters Proline Cubemass 300 Modbus RS485 A0026892 Switch-off value Switch-on value Conductive Non-conductive Process variable Status output Behavior of status output when Switch-on value = Switch-off value: • Process variable > Switch-on value: transistor is conductive • Process variable < Switch-off value: transistor is non-conductive...
Page 145
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description Use this function to enter the limit value for the switch-on value (process variable > switch-on value = closed, conductive). When using a hysteresis: Switch-on value > Switch-off value.
Page 146
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting Mass flow Assign status Navigation Expert → Output → PFS output 1 to n → Assign status (0485–1 to n) Prerequisite • The Switch option is selected in the Operating mode parameter (→ 126).
Page 147
Proline Cubemass 300 Modbus RS485 Description of device parameters Failure mode Navigation Expert → Output → PFS output 1 to n → Failure mode (0486–1 to n) Description Use this function to select a failsafe mode for the switch output in the event of a device alarm.
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting Additional information Selection No option (passive - negative) A0026693 Yes option (passive - positive) A0026692 3.5.3 "Relay output 1 to n" submenu Navigation Expert → Output → Relay output 1 to n ‣...
Page 149
Proline Cubemass 300 Modbus RS485 Description of device parameters Switch status → 154 Powerless relay → 154 Terminal no. Navigation Expert → Output → Relay output 1 to n → Terminal no. (0812–1 to n) Description Displays the terminal numbers used by the relay output module.
Page 150
Description of device parameters Proline Cubemass 300 Modbus RS485 • Limit Indicates if a specified limit value has been reached for the process variable. Is used to output diagnostic information relating to the process and to react to it appropriately at the system level.
Page 151
Proline Cubemass 300 Modbus RS485 Description of device parameters • NSVa • S&W volume flow • Water cut • Oil density • Water density • Oil mass flow • Water mass flow • Oil volume flow • Water vol. flow •...
Page 152
Description of device parameters Proline Cubemass 300 Modbus RS485 Assign status Navigation Expert → Output → Relay output 1 to n → Assign status (0805–1 to n) Prerequisite In the Relay outp.func. parameter (→ 149), the Digital Output option is selected.
Page 153
Proline Cubemass 300 Modbus RS485 Description of device parameters Switch-on value Navigation Expert → Output → Relay output 1 to n → Switch-on value (0810–1 to n) Prerequisite In the Relay outp.func. parameter (→ 149), the Limit option is selected.
Page 154
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Selection • Actual status In the event of a device alarm, faults are ignored and the current behavior of the input value is output by the relay output. The Actual status option behaves in the same way as the current input value.
Page 156
Description of device parameters Proline Cubemass 300 Modbus RS485 User interface • Not used • 24-25 (I/O 2) • 22-23 (I/O 3) Additional information "Not used" option The double pulse output does not use any terminal numbers. Signal mode ...
Page 157
Proline Cubemass 300 Modbus RS485 Description of device parameters Value per pulse Navigation Expert → Output → Double pulse out → Value per pulse (0983) Description Use this function to enter the value for the measured value that a pulse is equivalent to.
Page 158
Description of device parameters Proline Cubemass 300 Modbus RS485 Measuring mode Navigation Expert → Output → Double pulse out → Measuring mode (0984) Description Use this function to select the measuring mode for the double pulse output. Selection •...
Page 159
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description The dictates of safety render it advisable to ensure that the double pulse output shows a predefined behavior in the event of a device alarm. Selection • Actual value In the event of a device alarm, the double pulse output continues on the basis of the current flow measurement.
Changing the byte sequence in the host system often requires a extensive knowledge and significant programming efforts. Endress+Hauser introduced the Byte order parameter (→ 162) for this reason. This makes it possible to use the standard settings of the host system and change the byte sequence on the measuring device by trial and error.
Page 163
Proline Cubemass 300 Modbus RS485 Description of device parameters achieved by changing the byte sequence, the settings for the byte sequence of the host system must be adapted accordingly. Byte transmission sequence Byte addressing, i.e. the transmission sequence of the bytes, is not specified in the Modbus specification.
Description of device parameters Proline Cubemass 300 Modbus RS485 Telegram delay Navigation Expert → Communication → Modbus config. → Telegram delay (7146) Description Use this function to enter a delay time after which the measuring device replies to the request telegram of the Modbus master.
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Selection • Off The terminating resistor is disabled. • On The terminating resistor is enabled. For detailed information about enabling the terminating resistor, see the Operating Instructions for the device, "Enabling the terminating resistor" section→ 7 Fieldb.writ.acc.
Description of device parameters Proline Cubemass 300 Modbus RS485 Device ID Navigation Expert → Communication → Modbus info → Device ID (7153) Description Displays the device ID for identifying the measuring device. User interface 4-digit hexadecimal number Device revision Navigation ...
Proline Cubemass 300 Modbus RS485 Description of device parameters 3.6.4 "Web server" submenu Navigation Expert → Communication → Web server ‣ Web server Webserv.language (7221) → 167 MAC Address (7214) → 168 DHCP client (7212) → 168 →...
Page 168
Description of device parameters Proline Cubemass 300 Modbus RS485 MAC Address Navigation Expert → Communication → Web server → MAC Address (7214) Description Displays the MAC address of the measuring device. User interface Unique 12-digit character string comprising letters and numbers Factory setting Each measuring device is given an individual address.
Page 169
Proline Cubemass 300 Modbus RS485 Description of device parameters Subnet mask Navigation Expert → Communication → Web server → Subnet mask (7211) Description Displays the subnet mask. User entry 4 octet: 0 to 255 (in the particular octet) Factory setting 255.255.255.0...
Description of device parameters Proline Cubemass 300 Modbus RS485 Login page Navigation Expert → Communication → Web server → Login page (7273) Description Use this function to select the format of the login page. Selection • Without header •...
Page 171
Proline Cubemass 300 Modbus RS485 Description of device parameters Select antenna (2713) → 175 Connection state (2722) → 175 Rec.sig.strength (2721) → 175 WLAN IP address (2711) → 173 Gateway IP addr. (2719) → 176 IP address DNS (2720) →...
Page 172
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting – Network security Navigation Expert → Communication → WLAN settings → Network security (2705) Description Use this function to select the type of security for the WLAN interface.
Page 173
Proline Cubemass 300 Modbus RS485 Description of device parameters WLAN password Navigation Expert → Communication → WLAN settings → WLAN password (2716) Description Use this function to enter the WLAN password. User entry – Factory setting – WLAN IP address ...
Page 174
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting 255.255.255.0 WLAN passphrase Navigation Expert → Communication → WLAN settings → WLAN passphrase (2706) Prerequisite The WPA2-PSK option is selected in the Security type parameter (→ 172).
Page 175
Proline Cubemass 300 Modbus RS485 Description of device parameters WLAN channel Navigation Expert → Communication → WLAN settings → WLAN channel (2704) Description Use this function to enter the WLAN channel. User entry 1 to 11 Factory setting...
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting High Gateway IP addr. Navigation Expert → Communication → WLAN settings → Gateway IP addr. (2719) Description Use this function to enter the IP address of the gateway.
Proline Cubemass 300 Modbus RS485 Description of device parameters Selection • Cancel • Reset + totalize Factory setting Cancel Additional information Selection Options Description Cancel No action is executed and the user exits the parameter. Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow values previously totalized.
Page 178
Description of device parameters Proline Cubemass 300 Modbus RS485 • Targ.corr.vol.fl • Carr.corr.vol.fl • GSV flow • GSVa • NSV flow • NSVa • S&W volume flow • Oil mass flow • Water mass flow • Oil volume flow • Water vol. flow •...
Page 179
Proline Cubemass 300 Modbus RS485 Description of device parameters SI units US units Imperial units • cm³ • af • gal (imp) • dm³ • ft³ • Mgal (imp) • m³ • fl oz (us) • bbl (imp;oil) • ml •...
Page 180
Description of device parameters Proline Cubemass 300 Modbus RS485 Operation mode Navigation Expert → Application → Totalizer 1 to n → Operation mode (0908–1 to n) Prerequisite One of the following options is selected in the Assign variable parameter (→ 177) of the Totalizer 1 to n submenu: •...
Page 181
Proline Cubemass 300 Modbus RS485 Description of device parameters Factory setting Totalize Additional information Selection Options Description Totalize The totalizer is started or continues running. Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Description of device parameters Proline Cubemass 300 Modbus RS485 Failure mode Navigation Expert → Application → Totalizer 1 to n → Failure mode (0901–1 to n) Prerequisite One of the following options is selected in the Assign variable parameter (→ 177) of the Totalizer 1 to n submenu: •...
Page 184
Description of device parameters Proline Cubemass 300 Modbus RS485 Actual diagnos. Navigation Expert → Diagnostics → Actual diagnos. (0691) Prerequisite A diagnostic event has occurred. Description Displays the current diagnostic message. If two or more messages occur simultaneously, the message with the highest priority is shown on the display.
Page 185
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Display Via the local display: the time stamp and corrective measures referring to the cause of the diagnostic message can be accessed via the key. Example For the display format: F271 Main electronics...
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information User interface The maximum number of days is 9999, which is equivalent to 27 years. 3.8.1 "Diagnostic list" submenu Navigation Expert → Diagnostics → Diagnostic list ‣ Diagnostic list Diagnostics 1 (0692) →...
Page 187
Proline Cubemass 300 Modbus RS485 Description of device parameters User interface Days (d), hours (h), minutes (m) and seconds (s) Additional information Display The diagnostic message can be viewed via the Diagnostics 1 parameter (→ 186). Example For the display format:...
Page 188
Description of device parameters Proline Cubemass 300 Modbus RS485 Diagnostics 3 Navigation Expert → Diagnostics → Diagnostic list → Diagnostics 3 (0694) Description Displays the current diagnostics message with the third-highest priority. User interface Symbol for diagnostic behavior, diagnostic code and short message.
Page 189
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Display Via the local display: the time stamp and corrective measures referring to the cause of the diagnostic message can be accessed via the key. Examples For the display format: •...
Description of device parameters Proline Cubemass 300 Modbus RS485 Timestamp Navigation Expert → Diagnostics → Diagnostic list → Timestamp Description Displays the operating time when the diagnostic message with the fifth-highest priority occurred. User interface Days (d), hours (h), minutes (m) and seconds (s)
Page 191
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR Recommendation NE 107: • F = Failure • C = Function Check • S = Out of Specification •...
Description of device parameters Proline Cubemass 300 Modbus RS485 Event list Navigation Expert → Diagnostics → Event logbook → Event list Description Displays the history of event messages of the category selected in the Filter options parameter (→ 190).
Page 193
A maximum of 11-digit character string comprising letters and numbers. Additional information Description Uses of the serial number • To identify the measuring device quickly, e.g. when contacting Endress+Hauser. • To obtain specific information on the measuring device using the Device Viewer: www.endress.com/deviceviewer Endress+Hauser...
Page 194
The device features are not directly readable from the order code. Uses of the order code • To order an identical spare device. • To identify the device quickly and easily, e.g. when contacting Endress+Hauser. Endress+Hauser...
Page 195
Proline Cubemass 300 Modbus RS485 Description of device parameters Ext. order cd. 1 Navigation Expert → Diagnostics → Device info → Ext. order cd. 1 (0023) Description Displays the first part of the extended order code. On account of length restrictions, the extended order code is split into a maximum of 3 parameters.
Description of device parameters Proline Cubemass 300 Modbus RS485 Factory setting 2.02.00 Additional information Description This electronic nameplate stores a data record for device identification that includes more data than the nameplates attached to the outside of the device. 3.8.4 "Mainboard module"...
Description of device parameters Proline Cubemass 300 Modbus RS485 Bootloader rev. Navigation Expert → Diagnostics → Display module → Bootloader rev. (0073) Description Displays the bootloader revision of the software. User interface Positive integer 3.8.10 "Data logging" submenu Navigation ...
Page 203
Proline Cubemass 300 Modbus RS485 Description of device parameters Assign chan. 1 Navigation Expert → Diagnostics → Data logging → Assign chan. 1 (0851) Prerequisite The Extended HistoROM application package is available. The software options currently enabled are displayed in the SW option overv.
Page 204
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description A total of 1000 measured values can be logged. This means: • 1000 data points if 1 logging channel is used • 500 data points if 2 logging channels are used •...
Page 205
Proline Cubemass 300 Modbus RS485 Description of device parameters Assign chan. 4 Navigation Expert → Diagnostics → Data logging → Assign chan. 4 (0854) Prerequisite The Extended HistoROM application package is available. The software options currently enabled are displayed in the SW option overv.
Page 206
Description of device parameters Proline Cubemass 300 Modbus RS485 Clear logging Navigation Expert → Diagnostics → Data logging → Clear logging (0855) Prerequisite The Extended HistoROM application package is available. The software options currently enabled are displayed in the SW option overv.
Page 207
Proline Cubemass 300 Modbus RS485 Description of device parameters Factory setting Additional information Description Once measured value logging has been started with the Data log.control parameter (→ 207), the device does not save any data for the duration of the time delay entered.
Page 208
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Selection • Done Measured value logging has been performed and completed successfully. • Delay active Measured value logging has been started but the logging interval has not yet elapsed.
Page 212
Description of device parameters Proline Cubemass 300 Modbus RS485 "Electronic temp." submenu Navigation Expert → Diagnostics → Min/max val. → Electronic temp. ‣ Electronic temp. Minimum value → 212 Maximum value → 212 Maximum value Navigation Expert → Diagnostics → Min/max val. → Electronic temp. → Maximum value (6051)
Page 213
Proline Cubemass 300 Modbus RS485 Description of device parameters "Medium temp." submenu Navigation Expert → Diagnostics → Min/max val. → Medium temp. ‣ Medium temp. Minimum value (6109) → 213 Maximum value (6108) → 213 Minimum value Navigation ...
Page 214
Description of device parameters Proline Cubemass 300 Modbus RS485 Minimum value Navigation Expert → Diagnostics → Min/max val. → Carr. pipe temp. → Minimum value (6030) Prerequisite For the following order code "Application package", option EB "Heartbeat Verification + Monitoring"...
Page 215
Proline Cubemass 300 Modbus RS485 Description of device parameters Minimum value Navigation Expert → Diagnostics → Min/max val. → Oscil. frequency → Minimum value (6071) Description Displays the lowest previously measured oscillation frequency. User interface Signed floating-point number Maximum value Navigation ...
Page 216
Description of device parameters Proline Cubemass 300 Modbus RS485 User interface Signed floating-point number "Oscil. damping" submenu Navigation Expert → Diagnostics → Min/max val. → Oscil. damping ‣ Oscil. damping Minimum value (6122) → 216 → 216...
Proline Cubemass 300 Modbus RS485 Description of device parameters Minimum value Navigation Expert → Diagnostics → Min/max val. → Signal asymmetry → Minimum value (6015) Description Displays the lowest previously measured signal asymmetry. User interface Signed floating-point number Maximum value Navigation ...
Description of device parameters Proline Cubemass 300 Modbus RS485 3.8.13 "Simulation" submenu Navigation Expert → Diagnostics → Simulation ‣ Simulation Assign proc.var. (1810) → 219 Proc. var. value (1811) → 220 → 220 Status inp. sim. (1355) Signal level (1356) →...
Page 219
Proline Cubemass 300 Modbus RS485 Description of device parameters Assign proc.var. Navigation Expert → Diagnostics → Simulation → Assign proc.var. (1810) Description Use this function to select a process variable for the simulation process that is activated. The display alternates between the measured value and a diagnostic message of the "Function check"...
Page 220
Description of device parameters Proline Cubemass 300 Modbus RS485 Proc. var. value Navigation Expert → Diagnostics → Simulation → Proc. var. value (1811) Description Use this function to enter a simulation value for the selected process variable. Subsequent measured value processing and the signal output use this simulation value.
Page 221
Proline Cubemass 300 Modbus RS485 Description of device parameters Description Use this function to select the signal level for the simulation of the status input. In this way, users can verify the correct configuration of the status input and the correct function of upstream feed-in units.
Page 222
Description of device parameters Proline Cubemass 300 Modbus RS485 Curr.out. 1 to n sim. Navigation Expert → Diagnostics → Simulation → Curr.out. 1 to n sim. (0354–1 to n) Description Use this function to switch simulation of the current output on and off. The display alternates between the measured value and a diagnostic message of the "Function check"...
Page 223
Proline Cubemass 300 Modbus RS485 Description of device parameters Description Use this function to switch simulation of the frequency output on and off. The display alternates between the measured value and a diagnostic message of the "Function check" category (C) while simulation is in progress.
Page 224
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description The desired simulation value is defined in the Pulse value 1 to n parameter. Selection • Off Pulse simulation is switched off. The device is in normal measuring mode or another process variable is being simulated.
Page 225
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description The desired simulation value is defined in the Switch status 1 to n parameter. Selection • Off Switch simulation is switched off. The device is in normal measuring mode or another process variable is being simulated.
Page 226
Description of device parameters Proline Cubemass 300 Modbus RS485 Additional information Description The desired simulation value is defined in the Switch status 1 to n parameter. Selection • Off Relay simulation is switched off. The device is in normal measuring mode or another process variable is being simulated.
Page 227
Proline Cubemass 300 Modbus RS485 Description of device parameters Additional information Description The desired simulation value is defined in the Pulse value parameter (→ 227). Selection • Off Simulation of the double pulse output is switched off. The device is in normal measuring mode or another process variable is being simulated.
Page 228
Description of device parameters Proline Cubemass 300 Modbus RS485 Event category Navigation Expert → Diagnostics → Simulation → Event category (0738) Description Use this function to select the category of the diagnostic events that are displayed for the simulation in the Diag.
Proline Cubemass 300 Modbus RS485 Country-specific factory settings Country-specific factory settings SI units Not valid for USA and Canada. 4.1.1 System units Mass Mass flow kg/h Volume Volume flow Corrected volume Corrected volume flow Nl/h Density kg/l Reference density kg/Nl Temperature °C...
Country-specific factory settings Proline Cubemass 300 Modbus RS485 Nominal diameter [mm] [kg/p] 0.01 4.1.5 On value low flow cut off The switch-on point depends on the type of medium and the nominal diameter. Nominal diameter On-value for liquid [mm] [kg/h] 0.08...
Proline Cubemass 300 Modbus RS485 Country-specific factory settings Nominal diameter [in] [lb/min] ¹/₂₄ 0.15 ¹/₁₂ 0.75 ¹/₈ ¹/₄ 4.2.3 Output current span Current output 1 to n 4 to 20 mA US 4.2.4 Pulse value For detailed information about the pulse value for measuring devices for custody transfer, see the Special Documentation for the device →...
Explanation of abbreviated units Proline Cubemass 300 Modbus RS485 Explanation of abbreviated units SI units Process variable Units Explanation Density g/cm³, g/m³ Gram/volume unit kg/dm³, kg/l, kg/m³ Kilogram/volume unit SD4°C, SD15°C, SD20°C Specific density: The specific density is the ratio of the density of the fluid to the density of water at a water temperature of 4 °C (39 °F), 15 °C (59 °F), 20 °C (68 °F).
Page 233
Proline Cubemass 300 Modbus RS485 Explanation of abbreviated units Process variable Units Explanation Pressure psi a Pounds per square inch (absolute) psi g Pounds per square inch (gauge) Mass oz, lb, STon Ounce, pound, standard ton Mass flow oz/s, oz/min, oz/h, oz/d...
Explanation of abbreviated units Proline Cubemass 300 Modbus RS485 Imperial units Process variable Units Explanation Density lb/gal (imp), lb/bbl (imp;beer), Pound/volume unit lb/bbl (imp;oil) Corrected Sgal (imp) Standard gallon volume Correct.vol.flow Sgal/s (imp), Sgal/min (imp), Standard gallon/time unit Sgal/h (imp), Sgal/d (imp)
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Modbus RS485 Register Information Notes 6.1.1 Structure of the register information The individual parts of a parameter description are described in the following section: Navigation: navigation path to the parameter Parameter...
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Function code Access type Register in accordance with "Modbus Register in accordance with Applications Protocol Specification" "Modicon Modbus Protocol Reference Guide" Read XXXX 3XXXX Example: mass flow = 2007 Example: mass flow = 32007...
Page 239
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information ‣ Low flow cut off → 261 ‣ Partial pipe det → 262 ‣ Measurement mode → 262 Select medium (6062) → 262 Select gas type (6074) →...
Page 240
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 ‣ I/O config. → 264 I/O 1 to n terminals (3902–1 to n) → 264 I/O 1 to n info (3906–1 to n) → 264 I/O 1 to n type (3901–1 to n) →...
Page 241
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Assign curr. 1 to n (0359–1 to n) → 266 Current span (0353–1 to n) → 266 Fixed current (0365–1 to n) → 266 0/4 mA value (0367–1 to n) →...
Page 242
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Val. at min.freq (0476–1 to n) → 269 Val. at max.freq (0475–1 to n) → 269 Measuring mode (0479–1 to n) → 269 Damping out. 1 to n (0477–1 to n) →...
Page 243
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Assign diag. beh (0806–1 to n) → 272 Assign status (0805–1 to n) → 272 Switch-off value (0809–1 to n) → 272 Switch-off delay (0813–1 to n) → 272 Switch-on value (0810–1 to n)
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information "Def. access code" wizard Navigation: Expert → System → Administration → Def. access code Parameter Register Data type Access Selection / User entry / User interface → Def. access code 8677 to 8684 String Read / Write...
Page 256
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 "Input values" submenu "Current input 1 to n" submenu Navigation: Expert → Sensor → Measured val. → Input values → Current input 1 to n Parameter Register Data type Access Selection / User entry / User interface →...
Page 257
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information "Relay output 1 to n" submenu Navigation: Expert → Sensor → Measured val. → Output values → Relay output 1 to n Parameter Register Data type Access Selection / User entry / User interface → ...
Page 258
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Sensor → System units Parameter Register Data type Access Selection / User entry / User interface → Volume flow unit (0553) 2103 Integer Read / Write 0 = cm³/s 1 = cm³/min 2 = cm³/h 3 = cm³/d...
Page 259
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Sensor → System units Parameter Register Data type Access Selection / User entry / User interface → 71 = gal/d (imp) 72 = Mgal/s (imp) 73 = Mgal/min (imp) 74 = Mgal/h (imp) 75 = Mgal/d (imp) 76 = bbl/s (imp;beer)
Page 260
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Sensor → System units Parameter Register Data type Access Selection / User entry / User interface → Corr. vol. unit (0575) 2106 Integer Read / Write 100 = Nl 101 = Nm³ 102 = Sm³...
Page 261
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information "User-spec. units" submenu Navigation: Expert → Sensor → System units → User-spec. units Parameter Register Data type Access Selection / User entry / User interface → Mass text (0560) String Read / Write Max. 10 characters such as letters,...
Page 262
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 "Partial pipe det" submenu Navigation: Expert → Sensor → Process param. → Partial pipe det Parameter Register Data type Access Selection / User entry / User interface → Assign variable (1860) 5106 Integer Read / Write...
Page 263
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Sensor → External comp. Parameter Register Data type Access Selection / User entry / User interface → Temperature mode (6184) 5515 Integer Read / Write 0 = Internal value 1 = External value...
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information 6.3.4 "Input" submenu "Current input 1 to n" submenu Navigation: Expert → Input → Current input 1 to n Parameter Register Data type Access Selection / User entry / User interface → Terminal no. (1611–1 to n)
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 6.3.5 "Output" submenu "Curr.output 1 to n" submenu Navigation: Expert → Output → Curr.output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Terminal no. (0379–1 to n)
Page 267
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Output → Curr.output 1 to n Parameter Register Data type Access Selection / User entry / User interface → 20 mA value (0372–1 to n) 1: 5915 to 5916 Float Read / Write...
Page 268
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Output → PFS output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Assign pulse 1 to n (0460–1 to n) 1: 2461 Integer Read / Write...
Page 269
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Output → PFS output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Assign freq. (0478–1 to n) 1: 2614 Integer Read / Write 0 = Off...
Page 270
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Output → PFS output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Failure mode (0451–1 to n) 1: 2367 Integer Read / Write 0 = Actual value...
Page 271
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Output → PFS output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Assign status (0485–1 to n) 1: 3374 Integer Read / Write 0 = Low flow cut off...
Page 272
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Output → Relay output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Assign limit (0807–1 to n) 1: 8248 Integer Read / Write 1 = Mass flow...
Page 273
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Output → Relay output 1 to n Parameter Register Data type Access Selection / User entry / User interface → Switch status (0801–1 to n) 1: 3518 Integer Read 1 = Open...
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Output → Double pulse out Parameter Register Data type Access Selection / User entry / User interface → Pulse output (0987) 7041 to 7042 Float Read Positive floating-point number Invert outp.sig. (0993)
Page 275
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information "Modbus data map" submenu Navigation: Expert → Communication → Modbus data map Parameter Register Data type Access Selection / User entry / User interface → Scan list reg.0 to 15 (7114) 0: 5001 Integer Read / Write...
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Application → Totalizer 1 to n Parameter Register Data type Access Selection / User entry / User interface → Control Tot. 1 to n (0912–1 to n) 1: 2608 Integer Read / Write...
Page 280
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 "I/O module 2" submenu Navigation: Expert → Diagnostics → I/O module 2 Parameter Register Data type Access Selection / User entry / User interface → I/O 2 terminals (3902–2) 6542 Integer Read 0 = Not used...
Page 281
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information "Carr. pipe temp." submenu Navigation: Expert → Diagnostics → Min/max val. → Carr. pipe temp. Parameter Register Data type Access Selection / User entry / User interface → Minimum value (6030) 7533 to 7534...
Page 282
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 "Data logging" submenu Navigation: Expert → Diagnostics → Data logging Parameter Register Data type Access Selection / User entry / User interface → Assign chan. 1 (0851) 2445 Integer Read / Write 0 = Off 1 = Mass flow 2 = Volume flow 3 = Correct.vol.flow...
Page 283
Proline Cubemass 300 Modbus RS485 Modbus RS485 Register Information Navigation: Expert → Diagnostics → Data logging Parameter Register Data type Access Selection / User entry / User interface → Data log. status (0858) 5937 Integer Read 0 = Done 1 = Stopped 2 = Active...
Page 284
Modbus RS485 Register Information Proline Cubemass 300 Modbus RS485 Navigation: Expert → Diagnostics → Simulation Parameter Register Data type Access Selection / User entry / User interface → Value curr.out 1 to n (0355–1 to n) 1: 5995 to 5996 Float Read / Write 3.59 to 22.5 mA...
Proline Cubemass 300 Modbus RS485 Index Index 0 … 9 0/4 mA value (Parameter) ....105, 112 Backlight (Parameter) ......28 0% bargraph value 1 (Parameter) .
Page 287
Proline Cubemass 300 Modbus RS485 Index C0 to 5 (6022) ......101 Enter access code (0003) ....13 Calibration factor (6025) .
Page 288
Index Proline Cubemass 300 Modbus RS485 Last backup (2757) ..... . . 29 Operating mode Linear expansion coefficient (1817) ... 92 Pulse/frequency/switch output 1 to n (0469–...
Page 289
Proline Cubemass 300 Modbus RS485 Index Signal mode Current output 1 to n (0379–1 to n) ..109 Current input 1 to n (1610–1 to n) ..104 Pulse/frequency/switch output 1 to n (0492–...
Page 290
Index Proline Cubemass 300 Modbus RS485 WLAN subnet mask (2709) ....173 Function Zero point (6195) ......100 see Parameter Zero point adjustment control (6196) .
Page 291
Proline Cubemass 300 Modbus RS485 Index Measuring value at maximum frequency (Parameter) Pulse/frequency/switch output 1 to n (Submenu) 56, ........135 Measuring value at minimum frequency (Parameter) .
Page 293
Proline Cubemass 300 Modbus RS485 Index User volume text (Parameter) ....69 User-specific units (Submenu) ....68 Value 1 display (Parameter) .