NOTE
Verify that the Operating mode field displays READY before performing any further
actions.
8. At the Operation Panel, select
9. Perform the daily shutdown wash routine, then verify the operation.

Cleaning and replenishing the cuvette conditioner bottle

Materials required:
• Deionized water
• Cuvette conditioner
Time: 10 minutes
Analyzer mode: READY
NOTE
The cuvette conditioner bottle may be cleaned when it is refilled, but must be cleaned at
least once every 2 months.
Figure 5-20. Cuvette conditioner bottle
1. Open the filter cap at the front of the cuvette conditioner bottle (
tube with the filter.
2. Disconnect the cuvette conditioner bottle level sensor connector.
3. Turn the connector counter-clockwise and pull it out.
4. Remove the bottle.
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
5. Empty any remaining contents of the bottle.
6. Rinse the bottle with deionized water and drain well.
Maintenance
.
WASH
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1 ISE buffer bottle
2 RRV (Reaction) bath oil
3 Isotonic saline diluent bottle
4 Cuvette detergent bottle
5 Cell conditioner bottle
5
), and pull up the