Checking Lamp Energy - Siemens ADVIA 2400 Maintenance Procedures

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Figure 5-15. Reservoir levels Upper and Lower
c. Replace the reservoir cover. Do not over tighten.
4. Replace the lamp access cover.
NOTE
If adding coolant does not clear the lamp coolant warning, call your local technical
support provider or distributor.

Checking lamp energy

NOTES
Check the lamp energy after cleaning or replacing cuvettes, and after replacing the lamp.
When completing weekly maintenance, be sure to perform the procedures in the
following sequence:
1. Weekly wash
2. Lamp energy check (this procedure)
3. Cuvette blank measurement
Materials required:
No materials required
Time: 15 minutes
Analyzer mode: READY
Maintenance
1 Upper Mark (Green Line)
2 Lower Mark (Red Line)
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.