Siemens ADVIA 2400 Maintenance Procedures page 52

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

1 Probe
2 Spring Clips
3 Probe Wire Screw
4 Joint Connector
5 Probe Tubing
6 Probe Guide
Figure 5-32. DPP probe without cover
6. Using pliers, if necessary, gently loosen the probe joint connector (
back on the tubing (
7. Gently flex and pull back on the tubing (
CAUTION
Be careful not to damage the flare end or kink the tube.
8. Loosen but do not remove the locking screw (
9. Loosen but do not remove the probe wire screw
wire from the post (
CAUTION
Do not force the wire or bend excessively, to avoid breaking the wire off of the probe
body mount.
10. Securely hold the probe arm (
the side closest to the black wire (
an open, locked position.
CAUTION
There is some spring resistance when attempting to open the clips. Do
probe arm to swing side to side when opening the clips.
11. Loosen but do not remove the wire lock screw (
12. Remove the black wire (
other blue wire attached.
13. Gently lift the probe (
the probe arm (
156
7 Black Wire
8 Wire Lock Screw
9 Probe Arm
10 Probe Wire Post
11 Joint Holder (and Locking Screw)
) approximately 1 cm.
5
).
10
9
) and open the 2 spring clips (
) going to the probe (
7
7
) from the post coming from the probe (
1
) up through the probe guide (
9
).
5
) to remove it from the end of the probe body.
).
11
), then remove the orange probe
(3
2
) by grasping each at
) and gently raising each to
1
8
).
6
), then carefully remove it from
ADVIA 2400 Operator's Guide
4
), then slide it
allow the
not
1
), but leave the