Washing The Lines - Siemens ADVIA 2400 Maintenance Procedures

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

9. Select
10. When prompted, select

Washing the lines

1. Remove the buffer bottle with the deionized water and replace it with a bottle filled
with a solution of 475 mL of deionized water and 25 mL probe wash 3 solution.
2. At the ISE Operation window, in the Bufferprime area, enter
3. Select
4. When prompted, select
Rinsing the lines
1. Replace the probe wash 3 solution bottle with a bottle of deionized water.
2. In the Bufferprime area, enter
3. Remove the dummy electrode.
4. Reinstall the Na, K, and Cl electrodes.
5. In the Initialize area, select
6. Before reinstalling the buffer-solution bottle, thoroughly rinse the buffer bottle cap,
float switch, and tube with deionized water and dry completely.
7. Install the buffer bottle or replace it if the volume is low.
Priming and initializing the ISE module
1. At the ISE Operation window, in the Bufferprime area, enter
2. To prime the line with buffer, select
3. When the priming is finished, verify that the electrodes are not leaking.
4. At the ISE Operation window, select
5. Run 10 pooled serum samples, or do an ISE CV check.
6. Perform calibration and run controls.
164
Execute
to execute buffer prime.
Yes
.
Execute
Yes
to execute the buffer prime.
Execute
in the Times field, then select
50
.
, then select
Execute
Exit
, then select
Yes
in the Times field.
50
.
Execute
15
in the Times field.
.
Yes
.
ADVIA 2400 Operator's Guide