BioOptix™ 10
Section 1 Introduction
1.3.4 BioOptix 10 Rear Panel
Controls and
Connections
Figure 1-4 BioOptix 10 rear panel controls and
connections
Table 1-5 BioOptix 10 Rear Panel Controls and Connections
Item Name
1
Cooling Fan
2
External Control Connector
3
Fuses
4
AC Power Connector
1-6
1
Description
Ventilates the BioOptix 10 case and cools internal components.
Connects to the Foxy 200 Fraction Collector and computer.
Rear Panel Accessible Fuses.
Must always be replaced with same type fuses:
100/117 VAC — 3 amp "T" type fuse
234 VAC — 1.5 amp "T" type fuses
Internal fuses are not accessible to the user and should only be
replaced by a trained service technician.
Pump Drive Board
F101 — 4.0 amp "T" type fuse.
Power One PS
F1 — Not Serviceable.
Lamp Drive Board
F1 — .375 amp "T" type fuse.
Case Bottom Assy
F4 — Not Serviceable (T2 primary).
F5 — Not Serviceable (T2 secondary).
F6 — Not Serviceable (T2 secondary).
The Column Module has an internal 1.2 amp circuit breaker and a ther-
mal breaker that should only be reset by a trained service technician.
Connects to AC power source. A dedicated power source is recommended.
Mains power cord is the disconnect device.