A.1 Specifications
(Subject to change without notice.)
Ranges:
(Six ranges with AutoRanging on all ranges)
1:
0–1 part per million (ppm) O
2:
0–10 ppm O
3:
0–100 ppm O
4:
0–1000 ppm O
5:
0–1 % O
6:
0–25 % O
Accuracy:
±2% of full scale when calibrated and operated at the same constant
temperature and pressure.*
±5% of full scale over the operating temperature range once tempera-
ture equilibrium is reached.*
*
Additional 0.04 ppm on 0-1 ppm range, generated by internal scrub-
ber (which is necessary to maintain these specifications).
Response Time: 90 % of final reading in less than 61 seconds at 25 °C
Operating Temperature:0-50°C (32 -122 °F)
Analog Signal Outputs:
Digital Data Lines:
Alarms:
2
2
2
2
2
For percent of full scale indication:
For range identification:
One RS-232C serial interface for 2-way communications
with separate host computer—for remote monitoring and
control of all functions.
Five user-programmable absolute-reading alarm setpoints,
with Form C, SPDT relays (3A resistive).
TELEDYNE ANALYTICAL INSTRUMENTS
2
one 0–1 V dc non-isolated signal
one 4–20 mA dc isolated signal
one 0–1 V dc non-isolated signal
one 4–20 mA dc isolated signal