Installation Checklist - Teledyne 3160 Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

3 Installation
3 Installation
3 Installation
3 Installation
3 Installation
3.6.3 RS-232 Serial Digital Port
The RS-232 port is configured as data terminal equipment (DTE) and uses a 9-pin
D connector wired as follows:
1.
No connection
2.
Transmit Data
3.
Receive Data
4.
N/C
5.
Ground
6.
N/C
7.
RTC
8.
CTS
9.
N/C
Connecting the analyzer to a modem requires a PC to modem cable. When con-
necting the analyzer directly to a PC, terminal or other data terminal equipment, a
"null-modem" cable is required.
The following serial interface default parameters are used:
3.7

Installation Checklist

Have you:
Checked for leaks from any of the rear panel connections?
Checked the sample and span gas pressures (5–10 psi)?
Checked the instrument air pressure (70–80 psi)?
3-10
Defaults
Programmable Options
1200 Baud
1200 Baud
8 Bits
5, 6, 7, or 8 Bits
No Parity
Even, Odd, or None
1 Stop Bit
1 or 2 Stop Bits
TELEDYNE ANALYTICAL INSTRUMENTS

Advertisement

Table of Contents
loading

Table of Contents