1. Enter set-up from the Analyze mode by pressing
SETUP.
2. Press SET to set the clock functions.
(If only one cell block is installed, TEST and INSTL
will be omitted.)
3. Enter the authorization code by using the SELEC, UP
and DOWN keys, and press ACCPT.
4. Press MORE.
5. Press CLOCK to select clock functions.
6. Enter the function that you would like to change.
SET: Sets the time.
ZERO: Sets the time for the next zero and the inter-
SPAN: Sets the time for the next span and interval
ALRMS: Displays alarm times and dates.
TELEDYNE ANALYTICAL INSTRUMENTS
val times for automatic zeroing.
times for automatic spanning.