Hach ORBISPHERE K1200 Basic User Manual page 9

Hide thumbs Also See for ORBISPHERE K1200:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
1 Cable
2 Shield
3 Instrument
It is vitally important to ensure the shielding is pinched and secured between the two washers to ensure the
shielding attaches directly to the instrument housing as a ground. Failure to do this could cause damage to
the instrument, and for sensor cables give incorrect readings.
5. Reattach and tighten the cable gland nut.
6. Attach the wires to the corresponding terminal block connections.
Connection to mains power supply
Power supply connection (low voltage instruments)
For low voltage instruments (10-30 VDC), connection to the mains power supply is with a 8-pin
BINDER connector (supplied).
Note: The connectors are grooved to avoid an incorrect fitting to the instrument.
Connect the power cable to the connector as follows:
Figure 2 BINDER connector
Power supply connection (high voltage instruments)
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Electrocution hazard. Always remove power to the instrument before making electrical connections.
High voltage instruments (100-240 VAC) have a 4-pin male connector pre-wired internally with a
male BINDER connector ready for mains connection. A compatible female connector is supplied with
the instrument.
If this female connector was supplied with a mains power plug already pre-attached (cable part
numbers 33031, 33032, 33033 and 33034) then the female connector can be plugged directly into
4 Wire
5 O-ring
6 Washers
N O T I C E
Pin Connections:
1. Power 10-30 VDC
2. Ground
3. Ground
4. Ground
5. Not used
6. Power 10-30 VDC
7. Power 10-30 VDC
8. Earth
D A N G E R
D A N G E R
7 Gasket
8 Gland nut
Figure 3 Wiring side view
English 9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orbisphere 510

Table of Contents