Checking And Tightening Mast Pivot Bolt; Replacing The Swivel Bar - Invacare Low Profile RPL450-2 Owner's Operator And Maintenance Manual

Electric portable patient lift
Hide thumbs Also See for Low Profile RPL450-2:
Table of Contents

Advertisement

MAINTENANCE

Checking and Tightening Mast Pivot Bolt

NOTE: For this procedure, refer to FIGURE 7.3.
1. Lift up the back of the rubber boot and 
slide it off the mast along the boom.
2. Check that the bolt is through the 
bracket and the locknut is tight and 
secure.
3. If needed, do one or more of the 
following:
• Tighten locknut and back‐off the 
locknut 1/8 of a turn.
• Replace the locknut.
4. Reposition the rubber boot.

Replacing the Swivel Bar

After the first year of use, the hooks of the swivel bar and mounting brackets of the
boom should be inspected every six months to determine the extent of wear. If
these parts become worn, replacement must be made.
NOTE: For this procedure, refer to FIGURE 7.4 on page 39.
1. The swivel bar comes attached to the boom with bolt, nut, washers, and pinch guard.
NOTE: Washers should be placed against both sides of the boom mounting bracket to eliminate 
damage.
NOTE: The swivel bar has three hooks on both ends to attach the sling that supports the patient 
during lift.
2. Remove existing hardware and replace the swivel bar.
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
Electric Portable Patient Lift
Washer
Bolt
Mast
FIGURE 7.3 Mast Pivot Bolt
WARNING
CAUTION
38
Rubber Boot
Boom
Spacer
(if equipped)
Locknut
Bracket
Part No. 1145810