Invacare® ISA™ ISA™ COMPACT, ISA™ STANDARD, ISA™ PLUS, ISA™ XPLUS en Stand assist patient lift Service Manual DEALER: Keep this manual. The procedures in this manual MUST be performed by a qualified technician.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
CAUTION Find the user manual on Invacare’s website or contact Indicates a hazardous situation that could your Invacare representative. See addresses at the end result in minor or slight injury if it is not of this document.
Assembly of accessories might not be described in this service manual. Refer to the manual, delivered with the accessory. – Additional manuals can be ordered from Invacare. See addresses at the end of this document. IMPORTANT! Some replacement parts are only available as a kit.
Page 6
Invacare® ISA™ CBJ Care, CBJ1 or CBJ2 control unit with Mechanical emergency lowering detachable battery Lever for manual leg spreader 60121094-A...
If the service light is required to be set for a period shorter than 12 months, a special handset and further instructions If not otherwise stated in the specific instructions the are required. Contact Invacare to order this handset and following guidelines apply: instructions.
3.4 Service inspection checklist Model: REF / SN number: Comment Checkpoints for Invacare® ISA™ according to ISO 10535:2006 Annex B ü û General inspection No unauthorized modifications All labels present and readable. (see full list of labels in the user manual)
Page 9
Servicing Model: REF / SN number: Comment Checkpoints for Invacare® ISA™ according to ISO 10535:2006 Annex B ü û Functional tests Castors smoothly swivel and roll Application and release of castor brakes is working properly Base fully opens and closes smoothly...
Invacare® ISA™ 4 Troubleshooting 4.1 Identifying and repairing faults Symptoms Faults Solution Patient lift feels loose Mast/Base joint loose Check and tighten the Mast/ Base connection. See “Installing the Mast” in the user manual Mast/Boom joint loose Check and tighten the Mast/ Boom connection.
Page 11
Clear the back of the leg support adjusted leg support The adjustment mechanism is damaged Replace the leg support. See 5.16 Replacing the leg support, page 26 Contact your Invacare provider or representative if the above does not solve your problems. 60121094-A...