SECTION 3—OPERATION

Installing the CPR Plug

NOTE: For this procedure, refer to FIGURE 3.4.
1. Insert the rubber plug into the hose on the head air sack.
NOTE: Moisten the rubber plug if necessary to ease installation.
2. Extend the CPR flag outside the mattress assembly to hang down beside the bed
frame.

Installing the Therapy Pad

A patient should not be placed directly on the mattress without a therapy pad. A
loose (non fitted) cotton sheet between the patient and therapy pad is optional.
A tightly fitted cotton sheet may increase skin friction and shear.
1. Starting at the foot end of the bed, lay the therapy pad (quilted side down) on top of
the mattress.
2. Connect the therapy pad snaps to the corresponding snaps on the straps located on
the mattress (near the bed frame).

Installing the Power Cord

1. Plug the power cord into a wall outlet.
NOTE: Verify that air hoses and power cord are clear of moving bed components.
2. Operate all bed motorized functions through their full range.
NOTE: Check for pulling, interference or pinching, and correct any problems.
MicroAIR®3500S
Air Sacks
Rubber
Plug
Mattress
Assembly
CPR Tag
FIGURE 3.4 Installing the CPR Plug
WARNING
16
Part No 9692006