Ma60 series alternating pressure system alternating pressure on-demand low air loss system alternating pressure with low air loss system - bariatric (42 inch) alternating pressure with low air loss system - bariatric (48 inch) alternating pressure with lo

Advertisement

MA65 Alternating Pressure On-Demand Low Air Loss System
MA65B42 microAir Alternating Pressure with Low Air Loss System - Bariatric (42 inch)
MA65B48 microAir Alternating Pressure with Low Air Loss System - Bariatric (48 inch)
MA65RSR microAir Alternating Pressure with Low Air Loss System and Raised Side Rails
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Invacare®MA60 Series
MA60 Alternating Pressure System
User Manual
EN

Advertisement

Table of Contents
loading

Summary of Contents for Invacare MicroAir MA60

  • Page 1 MA65RSR microAir Alternating Pressure with Low Air Loss System and Raised Side Rails This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference. Invacare®MA60 Series MA60 Alternating Pressure System User Manual...
  • Page 2 © 2011 Invacare Corporation. All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
  • Page 3: General

    Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or both. IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided. Gives useful tips, recommendations and information for efficient, trouble-free use.  Part No 1148137 2 GENERAL Invacare®MA60 Series...
  • Page 4: Limited Warranty

    MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare or a dealer. This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner. Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person.
  • Page 5: Table Of Contents

    Transferring Patient From/To a Gurney... 28 Transferring Patient From/To a Wheelchair ... 29 Preparing for CPR Procedure ... 29 About Power Outage and Transportation... 30 MAINTENANCE AND TROUBLESHOOTING Cleaning the System ... 31 Troubleshooting... 34 Part No 1148137 CONTENTS Invacare®MA60 Series...
  • Page 6: Overview

    Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions. Monitor patient frequently. Read and understand the Owner’s/Operator’s Manual prior to using this equipment. Invacare product manuals are available at www.invacare.com or your dealer.
  • Page 7: Typical Product Parameters

    INPUT VOLTAGE AC: INPUT FREQUENCY: CURRENT: MAXIMUM POWER CONSUMPTION: CIRCUIT PROTECTION: MODE OF OPERATION: Part No 1148137 MA60/MA65 110 V 60 Hz 30 ± 10 W Dual fused, 250 V, 1 A fast blow fuses Continuous 4 OVERVIEW Invacare®MA60 Series...
  • Page 8 MAXIMUM FLOW PRESSURE: MAXIMUM FLOW TIMER: SUPPORT SURFACE INFLATION TIME: PATIENT COMFORT CONTROL PRESSURES SOFT PRESSURE: FIRM PRESSURE: CYCLE TIME: PATIENT CONTACT: Invacare®MA60 Series MA60/MA65 350 lbs 650 lbs 1000 lbs 50 ± 15 LPM 35 ± 5 mmHg 30 minutes 5 -10 minutes 8 ±...
  • Page 9 Two ¼” Flow Couplings Three ¼” Flow Couplings 1 piece/box None MA60/MA65 50° - 95° F 30% - 75% Non-Condensing 70 - 106 kPa -40° - 158° F 10% - 100% 50 - 106 kPa 4 OVERVIEW MA65 Invacare®MA60 Series...
  • Page 10: General Guidelines

    Invacare and are not recommended for use with Invacare products. NOTICE THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE. Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact Invacare/Carrier for further instruction. Contraindications WARNING Always consult the patient’s physician before using the MA60, MA65, MA65B42, MA65B48, and MA65RSR systems.
  • Page 11 Check that air hoses and power cord are clear of the moving bed components before placing a patient on the bed. Operate all motorized functions through their full range of motion to be certain that there is no pulling, interference or pinching. Part No 1148137 5 GENERAL GUIDELINES Invacare®MA60 Series...
  • Page 12: Entrapment May Occur

    Visit the FDA website at http://www.fda.gov to learn about the risks of entrapment. Review “A Guide to Bed Safety”, published by the Hospital Bed Safety Workgroup, located at www.invacare.com. Use the link located under each bed rail product entry to access this bed safety guide.
  • Page 13 Oxygen There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type. The oxygen tent should NOT extend below mattress support level. Part No 1148137 5 GENERAL GUIDELINES Invacare®MA60 Series...
  • Page 14: Electrical

    The control unit MUST be kept away from all heat sources and radiators during operation. Connect the equipment to properly grounded three prong wall outlet using 10-14 ft hospital grade power cord provided with the product. Grounding reliability depends upon a properly grounded receptacle (3-prong). Invacare®MA60 Series Part No 1148137...
  • Page 15: Operation

    Otherwise damage to the mattress will occur when the head and foot ends are raised. Head End of Bed Deck DO NOT Strap Here Part No 1148137 Center of Bed Frame Foot End of Bed Deck 6 OPERATION Invacare®MA60 Series...
  • Page 16 Head End - Head End Bed Deck • Foot End - Foot End Bed Deck • Center - Center of the Bed Frame Head End Buckle Straps Invacare®MA60 Series Hose FIGURE 1 Installing the Powered Mattress Powered Mattress Foot End Part No 1148137...
  • Page 17: Installing The Bed Side Rails

    Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions. Monitor patient frequently. Read and understand the Owner’s/Operator’s Manual prior to using this equipment. Invacare product manuals are available at www.invacare.com or your dealer.
  • Page 18: Installing The Control Unit

    Insert the hose connectors into the control unit connectors. Push the hose connectors into the control connectors until an audible click is heard.  The audible click indicates that the hose connectors are properly engaged with the control connectors. Invacare®MA60 Series Part No 1148137...
  • Page 19 6 OPERATION Control Unit Connectors Hose Connector Hose FIGURE 2 Connecting the Hose Part No 1148137 Invacare®MA60 Series...
  • Page 20: Connecting The Power Cord

    Plug the other end of the plug into a properly grounded outlet on the wall.  Once the unit is plugged in, an Amber led on the control unit is lit indicating that the system is in stand by mode. Invacare®MA60 Series Part No 1148137...
  • Page 21: Using The Front Panel

    Firm Button FIGURE 3 Using the Front Panel Fowler Button Power Fail LED Lock/Alarm Button Max Inflate Button Power Fail LED Fowler Button Mode Button (Max Inflate/ Low Air Loss) Invacare®MA60 Series 6 OPERATION Low Pressure LED Low Pressure LED Lock/Alarm Button...
  • Page 22 50% of the selected pressure setting. Ensure the Standby LED is lit. Press and hold the Alternating Pressure ( Invacare®MA60 Series ) to select between one of four alternating pressure times - 5, 10, 15 and 20 minutes. ) and Fowler ( ) buttons.
  • Page 23 If the patient’s weight to height ratio is above average, increase the pressure setting by approximately 20%. Part No 1148137 ) buttons to toggle between the two settings. ) or Soft ( ) button. 6 OPERATION Invacare®MA60 Series...
  • Page 24 MA65 - Press and hold the Mode and Fowler buttons.  The Fowler LED is lit and L displays in the LCD Display. Raise the bed to 45°.  The control unit will beep and will return to Standby. Invacare®MA60 Series Part No 1148137...
  • Page 25 Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the previous setting. Part No 1148137 ) to select the max inflate mode or the low air loss mode. 6 OPERATION Invacare®MA60 Series...
  • Page 26 In the event that the mattress hose disconnects, an alarm sounds and Low Pressure LED flashes AMBER. Once the low pressure problem is fixed, the control unit resumes operation in the previously set mode. Invacare®MA60 Series ) to select one of the following modes:...
  • Page 27: Powering Up The System

    Wait five minutes for the mattress pressure to stabilize. Once the mattress inflates to its normal size, set the comfort pressure to the desired comfort level. Wait five minutes for the mattress pressure to stabilize. Part No 1148137 6 OPERATION Invacare®MA60 Series...
  • Page 28: Transferring Patient From/To A Gurney

    When the mattress has reached maximum firmness, perform one of the following: • Bed to Gurney Transfer - Slide the patient onto the gurney. • Gurney to Bed Transfer - Slide the patient onto the bed. Invacare®MA60 Series Part No 1148137...
  • Page 29: Transferring Patient From/To A Wheelchair

    6.11 Preparing for CPR Procedure Press and hold the tabs on hose connector while pulling the hose from the control unit. Disconnect the RED CPR connector located on the side of the mattress. Part No 1148137 6 OPERATION Invacare®MA60 Series...
  • Page 30: About Power Outage And Transportation

    If the hoses remain connected to the control unit, the mattress retains air during a power outage, during transportation, when the control unit is unplugged or when the control unit is turned Off. If the hose becomes disconnected or damaged and the mattress deflates, the mattress has a 2-inch foam pad to provide patient support. Invacare®MA60 Series RED CPR Connector FIGURE 4 Preparing for CPR Procedure...
  • Page 31: Maintenance And Troubleshooting

    Scrub the area with freshly prepared effective phenolic detergent disinfectant solution. • If blood is not present, remove any soil from the cover with paper towels.  If soiled, the cover should be removed, cleaned and decontaminated. Part No 1148137 7 MAINTENANCE AND TROUBLESHOOTING Invacare®MA60 Series...
  • Page 32 DO NOT overload the machine. DO NOT use chlorine bleach because it may damage the fabric coating. High air temperatures will damage the fabrics and void the Invacare warranty. Place the cover in a washing machine. D. Wash with warm water (below 120°F).
  • Page 33 Disconnect the air hose connector(s) from the control unit and allow air to vent from the mattress. Gently roll up the mattress with minimal handling and agitation. Ensure the cover surface is inside the roll. Store the unit, keeping the mattress with the control unit. Part No 1148137 7 MAINTENANCE AND TROUBLESHOOTING Invacare®MA60 Series...
  • Page 34: Troubleshooting

    Control unit off Power cord disconnected No power in the power source Power outage Blown fuse Invacare®MA60 Series CAUSE Connect hose connectors and lock them in place Unkink hose or replace split hose Replace leaking air cushions or overlay pad...
  • Page 35 NOTES NOTES Part No 1148137 Invacare®MA60 Series...
  • Page 36 Part No 1148137 Rev D- 1/11 Invacare Corporation Canada One Invacare Way 570 Matheson Blvd E Elyria, Ohio USA Unit 8 44036-2125 Mississauga Ontario 800-333-6900 L4Z 4G4 Canada 800-668-5324 www.invacare.com...