II 3058-4
Select TOTAL CYCLE COUNTER ..................................................press ENTER
Select YES ........................................................................................press ENTER
Press BACKSPACE 3 times
Select QUIT ML300 SERVICE MODE ...........................................press ENTER
Select Quit the program .............................................................press ENTER

6. SWITCH SETTING OF S1-1 AND S1-2 ON PCB A1

S1-1
Normally S1-1 should be set to OFF (down). In this case some errors can only be reset with
the LAP TOP (see the SERVICE MANUAL SUPPLEMENT SN>2000).
If S1-1 is set to ON (up) the following options are enabled:
1.
The OPERATOR can reset the XML 300 after every malfunction just by pressing the
left-hand BUTTON. He can then try to start another cycle.
This RESET cannot solve a real hardware problem or a film jam when it tries to recover from the
problem. In such a case the problem will occur again in the next cycle. This means an exposed
!
Caution
FILM would become lost. For this reason set S1-1 only to ON if the CUSTOMER is aware of this
risk and if he is willing to accept it.
2.
After selecting "FUNCTION" on the OPERATOR CONTROL PANEL the following
selection is displayed:
TYPE2
When S1-1 is ON (up) an altered selection is displayed:
TYPE2
SYSTEM
SYSTEM
contains the following 4 options:
TIME
This gives date and time.
HISTO
Note
THIS INFORMATION CAN ONLY BE ANALYSED IN THE FACTORY.
KODAK AG Stuttgart
USAGE
USAGE
This gives information on the communication between the MAIN PROCESSOR
PCB A1 and the SLAVE PROCESSORS PCB's A3/1, A3/2 and the 270RA
PROCESSOR.
LANG
CLEAR
LANG
CLEAR
10-5