Temperature Verification Test; Agilent U1602A/U1604A Handheld Digital Oscilloscope Test Record - Agilent Technologies U1602A User's And Service Manual

Handheld digital oscilloscopes
Hide thumbs Also See for U1602A:
Table of Contents

Advertisement

Agilent U1602A/U1604A Handheld Digital Oscilloscope Test Record

Serial No.
Test Interval
Recommended Next Testing
Oscilloscope Performance Verification Tests
Voltage Measurement Accuracy
Volt/div
Power Supply Setting
100 V/div
250 V
50 V/div
250 V
20 V/div
140 V
10 V/div
70 V
5 V/div
35 V
2 V/div
14 V
1 V/div
7 V
0.5 V/div
3.5 V
0.2 V/div
1.4 V
0.1 V/div
700 mV
50 mV/div
350 mV
20 mV/div
140 mV
10 mV/div
70 mV
5 mV/div
35 mV
Bandwidth
Model
Test Limits
U1602A
±3 dB at 20 MHz
U1604A
±3 dB at 40 MHz
Horizontal Time Base Accuracy
Model
Calibrator Setting
U1602A
Frequency: 10 kHz
Period: 100
Frequency: 1 MHz
Period: 1
Frequency: 20 MHz
Period: 50 ns
Agilent U1602A/U1604A User's and Service Guide
Test Limits
226 V to 274 V
238 V to 262 V
135.2 V to 144.8 V
67.6 V to 72.4 V
33.8 V to 36.2 V
13.52 V to 14.48 V
6.76 V to 7.24 V
3.38 V to 3.62 V
1.352 V to 1.448 V
676 mV to 724 mV
338 mV to 362 mV
132 mV to 148 mV
66 mV to 74 mV
33 mV to 37 mV
Test Limits
9.63 kHz to 10.40 kHz
s
96.20
m
962 kHz to 1.04 MHz
s
0.96
m
18.69 MHz to 21.51 MHz
46.5 ns to 53.5 ns
Test by
Work Order No.
Temperature
Channel 1
s to 103.8
s
m
m
s to 1.04
s
m
m
Performance Tests
Channel 2
Channel 1
Channel 2
Results
7
179