Invacare® ISA™

5.9 Charging the battery

IMPORTANT!
– Make sure the emergency stop is not activated
while charging the battery.
– Make sure that charging takes place in a room
with good air ventilation.
– Electrical functions are not working when the
lift is connected to power supply.
– Do not attempt to use the lift if the battery
housing is damaged.
– Replace a damaged battery housing before
further use.
– Do not move the lift without unplugging from
the power outlet.
It is recommended to charge the battery daily to ensure
optimal use of the lift and prolong the life of the battery.
Furthermore, it is recommended to charge the battery
before first use.

5.9.1 CBJ Home control unit

The control unit is equipped with a sound signal. A beep
indicates that the battery has low capacity, but lowering the
patient is still possible. It is recommended to charge the
batteries as soon as the sound signal is heard.
1. Plug the power cord C into a power outlet.
The battery will charge in approximately 4
hours. The charger stops automatically when the
batteries are fully charged.
The upper yellow diode A will flash during
charging, and switch to continuous light when
fully charged.
The lower green diode B will light continuously
while the control unit is connected to the
mains, and lights up when any button on the
hand control is pressed or when the electric
emergency lowering is activated.
2. Disconnect the power cord from the power outlet after
the battery has been fully charged.

5.9.2 CBJ Care, CBJ1, CBJ2 control unit

The control unit is equipped with a sound signal. A beep
indicates that the battery has low capacity, but lowering the
16
patient is still possible. It is recommended to charge the
batteries as soon as the sound signal is heard.
1. Plug the power cord A into a power outlet.
The battery will charge in approximately 4
hours. The charger stops automatically when the
batteries are fully charged.
The right yellow diode B will light continuously
during charging, and switch off when fully
charged.
The left green diode C will light continuously
while the control unit is connected to the mains.
2. Disconnect the power cord A from the power outlet
after the battery has been fully charged.
Battery Indicator
The control unit may be equipped with a battery indicator
D, indicating the remaining battery capacity.
CBJ Care
Battery
Battery
Indicator
State
Type
Full
Charge
Partial
Charge
Low
Charge
Low
Charge
(LED
flashing)
CBJ1 and CBJ2 with LCD
Battery
Battery
Indicator
State
Type
Full
Charge
(100%)
Partial
Charge
(75%)
Partial
Charge
(50%)
Description
The battery is OK — no need
for charging (100–50%). The
top LED is GREEN.
The battery needs to be
charged (50–25%). The middle
LED is YELLOW.
The battery needs to be
charged (Less than 25%).
Acoustic signal when a button
is pressed. The bottom LED is
YELLOW.
The battery needs to be
charged.
Some of the functionality of
the lift is lost and it is only
possible to lower the boom.
Description
The battery is OK — no need
for charging (100%).
The battery is OK — no need
for charging (75%).
The battery needs to be
charged (50%).
1654371-A