User Interfaces

Depending on the used interface, the available tests and the screens/reports may vary.
N O T E
Preferred tool should be the Agilent Diagnostic Software, see
Software"
Screenshots used within these procedures are based on the Agilent ChemStation.
In future, a user interface may not show the Diagnostics/Tests anymore. Then the Agilent
Diagnostic Software must be used instead.
The Agilent ChemStation may not include any maintenance/test functions.
Table 18
Test
Selftest
Filter
Slit
D/A Converter
Test Chromatogram
Wavelength Calibration
Lamp Intensity
Holmium
Cell
Dark Current
*
Agilent 1260 Infinity DAD and MWD User Manual
on page 122.
Test Functions avaible vs. User Interface
Diagnostic Software
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
requires a command via command line
Troubleshooting and Diagnostics

User Interfaces

"Agilent Lab Advisor
ChemStation
Instant Pilot G4208A
Yes
No
Yes
No
Yes
No
Yes
No
Yes (*)
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
6
121