The plugs on power module cannot be wrongly
connected because all plugs have a different size
and only fit in one socket.
•
10 mm wrench
•
17 mm wrench
•
Rubber hammer
•
Phillips screwdriver
•
Supporting wooden block
•
To adapt the drive program: programming
software or hand programming device and
controls system installation manual, available
from Invacare.
1. Secure mobility device against rolling away.
2. Place supporting wooden block under frames at
battery holder height.
3. Remove seat.
4. Remove shroud.
5.
Disconnect battery cable (1).
6. Remove rear wheel fixing screws.
7. Remove rear wheels.
8. Remove drive unit. See 5.10.1 Replacing drive unit,
page 30.
9. Remove screws (1) on power module (2).
•
Rhino1:
Fig. 5-7 DS112KB02/DS162KD01
•
Rhino 2:
1603798-D
Fig. 5-8 DS120
Fig. 5-9 DS160
10.
Replace power module (2).
11. Install parts in reverse order.
12. Load drive program onto power module. See controls
software description.
13. Test all functions (trial run).
5.6 Shrouds
5.6.1 Removing shroud
CAUTION!
Risk of damage to shroud
– Do not pull shroud straight up to remove.
– Follow instructions below.
1.
Place one hand at the front, the other at the rear of
the shroud A.
2. Lift shroud up from front.
3.
Place hands on left and right side of shroud.
4. Twist and lift to remove shroud completely.
5.6.2 Replacing front shroud
•
10 mm socket wrench
•
Phillips screwdriver
Service