Power plugs and cords
The Agilent ICP-OES is supplied with a power cord set appropriate
for the country from which the order originates. See Appendix A for
illustrations of the power cords available.
If necessary, replace the power cord only with a cord equivalent to
the one specified (See Appendix A).
If an instrument is being ordered from one location, but is to be installed in
CAUTION
another location with different electrical power characteristics, this must be
noted on the order. A special note must also be made if the electrical power at
the site is different from the standard electrical power in that country.
WARNING
Make sure the power cords supplied with the Agilent ICP-OES are
appropriate for your country and site before using them.
After the electrical requirements have been met, check the checklist box: Specified
electrical supply and power outlets installed.
Gas Supplies
Liquid or gaseous argon and nitrogen may be used with Agilent
ICP-OES spectrometer systems. Agilent recommends the use of liquid
gases, which are purer, more convenient and cheaper per unit
volume. Contact your local authorities and gas supplier for more
detailed information on storage requirements for all required gases
and also boil-off rates for local types of or Portable Liquid Cylinders
(PLC).
Table 6. Gas supply requirements
Component
Plasma, nebulizer, optics interface purge
Agilent 5100 and 5110 ICP-OES Site Preparation Guide
Laboratory Facilities
Gas
Argon
27