6890A Gc; Follow The Instructions Below To Properly Wire The G2912A Controller. See "Procedure 8. Install The G2912A Als Controller" Instructions For The G2912A Controller Including The Power Supply Cord; Power Cords Not Shown; Connect The Injector(S) To The Controller - Agilent Technologies agilent 7683B Installation, Operation And Maintenance Manual

Automatic liquid sampler
Table of Contents

Advertisement

6890A GC

Follow the instructions below to properly wire the G2912A controller. See
"Procedure 8. Install the G2912A ALS Controller"
instructions for the G2912A controller including the power supply cord.
Figure 18

1 Connect the injector(s) to the controller.

2 Connect the tray to the controller with G2614-60610 cable.
3 Connect the controller to the GC with a G1530-60930 cable and a
4 Connect the Bar Code Reader to the tray, if desired.
5 Connect the power cords.
Installation, Operation, and Maintenance
Part 1, Installation

G2912A Controller

Front
Back
Injector
Injector
Inj1
Inj2
Front
Back
Injectors

Power cords not shown

Cables for 6890A GC
G1530-60600 cable.
6890
5890
Tray
Remote
Remote
G2614-60610
Tray
G1530-60930
Bar Code
Reader
Installation

on page 55 for installation

RS-232
G1530-60600
Sampler
6890A GC
Remote
2
47