7 Error and Troubleshooting

Symptom
No display when power On
Display "Connect Error"
Display "DATA Error"
Press "START" but no response
Testing record can not be saved
New ID can not be added
Printing unreadable code
Possible causes
incorrect specification of the AC
Adapter
Battery installation error
Low or dead battery power
Connection cable loss with the
base
Back-Pressure
mode
incorrect
Low battery charge
Back Pressure Plate opened On
Over-limit Cycle quantity in
current ID
ID over-limit
Current ID is without "Cycle"
Unmatched printer
Wrong baud rate setting
23
Corrective action
Check and refer to Chapter4.1
in Manuel
Check and refer to Chapter4.1
in Manuel
Replace the batteries
Reseat the connect cable
using
Refer to Chapter 5.4 in Manuel
Replace new batteries
Shut off the plate
Add new ID or delete some
data
Delete some ID
Using the current ID
Using our special optional
printer supplied by us.
Reset referring to the Printer
Manuel