A.7 Optical Output
CA UT I O N
The attenuator is supplied with either a straight contact connector or
an angled contact connector (Option 201). Make sure that you only use
the correct cables with your chosen output. See "Connector Interfaces
and Other Accessories" on page 158 for further details on connector
interfaces and accessories.
NO T E
Depending on the attenuation setting, it can take up to 3 seconds for the
output to be disabled (typically delay, 1 second).
A.8 GPIB Interface
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Installation

Optical Output

coupling ratio, and its wavelength dependence, for the Monitor
Output yourself.

Disabling the Optical Output

If the optical output is enabled (that is, the green LED is lit), you
can disable it by pressing E
You can connect your GPIB interface into a star network, a linear
network, or a combination star and linear network. The limitations
imposed on this network are as follows:
• The total cable length cannot exceed 20 meters
• The maximum cable length per device is 2 meters
• No more than 15 devices may be interconnected on one bus.