1.1 Using the Attenuator

NO T E
Before using the instrument, you should make sure that it is properly
warmed up. The instrument is properly warmed up when it has been
switched on for a minimum of 45 minutes. Failure to do this can cause
errors of up to 0.04dB in the attenuation.
Figure 1-1
The Attenuator Keys
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Getting Started
Using the Attenuator
Set the attenuation of the filter using A
(wavelength), and C
(calibration factor).
AL
The attenuation factor and the calibration factor set the position of
the filter. The calibration factor allows you to offset the value of the
attenuation factor.
Att(dB) = Cal(dB) + Attenuation
In addition, you can use D
ISP
attenuation factor to the calibration factor.