Sre; Table 8-7 The Service Request Enable Register - Agilent Technologies 8156A Operating And Programming Manual

Attentuator
Table of Contents

Advertisement

Table 8-7
The Service Request Enable Register
BITS
7
6
5
4
3
2
1
0
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Remote Commands
The Common Commands
Definition
The instrument setting is stored in RAM. You
can store settings in locations 1-9. The scope of
the saved setting is identical with the scope of
the standard setting described in "*RST" on
page 99.
Example
OUTPUT 728;"*SAV 3"

*SRE

Syntax
*SRE <wsp> <value>
0 ≤ value ≤ 255
Definition
The service request enable command sets bits
in the service request enable register that
enable the corresponding status byte register
bits.
The register is cleared:
At power-on
By sending a value of zero.
The register is not changed by the *RST and
*CLS commands.
MNEMONICS
Operation Status
Request Status
Event Status Byte
Message Available
Questionable Status
Not used
Not used
Not used
BIT VALUE
128
64
32
16
8
0
0
0