Installing/mounting
5.1 Basic safety notes
5.1.1.2

Avoid

● Proximity to high voltage or current wiring, high voltage or current contacts, and to variable
frequency motor speed controllers
● Exposure to direct sunlight. (Provide a sun shield to avoid direct sunlight.)
● Interference to sound path from obstructions (such as rough walls, seams, rungs, etc.), or
from fill path
20
Customer-supplied flange
Ambient temperature surrounding enclosure –40 to +80 °C (–40 to +176 °F)
Device tag (optional)
Pipe
Rungs
Seams
Compact Operating Instructions, 10/2018, A5E42673704-AB
SITRANS Probe LU240 with mA/HART