Section 1 Specifications - Hach BODTrak II User Manual

Hide thumbs Also See for BODTrak II:
Table of Contents

Advertisement

Section 1
Specifications are subject to change without notice.
General
Range
Dimensions
External power supply
Capacity
Shipping weight
Operating temperature
Storage temperature
Method performance specifications
Precision
Drift
Resolution
Certification
Hach Company certifies this instrument was tested thoroughly, inspected and found to meet
its published specifications when it was shipped from the factory. The BODTrak II has been
tested and is certified as indicated to the following instrumentation standards:
FCC Part 15, Sub-Part B, Class A Limits: Supporting test records by Intellistor, certified
compliance by Hach Company
Canadian Interference-Causing Equipment Regulation, ICES-003, Class A: Supporting
test records by Intellistor, certified compliance by Hach Company
EN 55011/CISPR 11(EMI) "B" Limits per 89/336/EEC EMC: Supporting test records by
Intellistor, certified compliance by Hach Company
EN 50082-1 (Immunity) per 89/336/EEC EMC: Supporting test records by Hach Company,
certified compliance by Hach Company. Standards include:
IEC 801-2 and EN 61000-4-2 (ESD)
IEC 801-3 and EN V50140 (RF & EM Field)
IEC 801-4 and EN 61000-4-4 (Fast Transient)
EN 61000-4-5 (Sur
Warranty: US 1 year; EU 2 year
Specifications
Table 1 Specifications
Selectable, 0 to 35, 0 to 70, 0 to 350, 0 to 700 mg/L
28.9 x 26 x 9.8 cm (11
Input: 110 to 240 V, 50/60 Hz, Output: 24 V, UL CSA, and
TUV approved
Six 492 mL bottles
4 kg (8.8 lb)
20 ºC (68 ºF)
0 to 40 ºC (104 ºF)
On a standard containing 150 mg/L each of glucose and
glutamic acid, a single analyst using 6 BODTrak™ II
instruments and testing 44 samples obtained a mean of
235 mg/L BOD with a 95% confidence limit of distribution
of 11 mg/L or a range of
224 to 246 mg/L BOD.
Less than 3 mg/L BOD in 5 days
1 mg/L BOD
Table 2 Certification
ge)
3
7
/
x 10 ¼ x 3
/
inches)
8
8
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents