Hach QbD1200 User Manual page 15

Hide thumbs Also See for QbD1200:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
Set the report settings
View, add or change the report templates. Refer to
reports and data.
1. Push Instrument Setup>Report Settings
2. Select an option.
Option
Description
Add
Makes a new report template. Configure the header and footer format and content.
Delete
Erases a report template.
Set default
Sets an existing report template as the default template.
Edit
Changes an existing template.
3. Push Save or Cancel.
Configure the measurement settings
1. Push Instrument Setup>Measurement Settings.
2. Select an option.
Option
Description
Default
Sets the default create tray wizard settings: the number of sample locations (1 to 64), sample
Settings
replicates (1 to 5), standard positions, default labels (date/time, ID individually, ID builder) and
sample views (tray or table).
Process
Sets the default sample process settings.
Settings
• Auto Ranging—Sets the instrument to adjust the measurement range automatically. Do not
select this option if the sample is less than 4 ppm.
• Conditioning—Sets the instrument to wait for the lamp temperature to increase before a
measurement is done to get accurate results. Conditioning only occurs on the first replicate of
the first sample on the tray. The manufacturer recommends that this option is selected.
• Homogenizing—Sets the autosampler to homogenize the sample before a measurement is
done. Only select this option when an autosampler is used.
• Dilution factor—Select this option if the sample is diluted before analysis. Select to prompt
the user to enter a dilution factor before a measurement is done. The instrument adjusts the
results to include the dilution factor.
Results
Sets the default display settings for the results.
Display
• Measurement Units—Sets the results to show in units of ppm or ppb. Select the "Use SI
Settings
Units" box to show the results in mg/L or µg/L.
• Pressure Units—Sets the internal air pressure of the instrument to show in units of kPa or
mbar.
• TIC Analysis—Sets the approximate TIC (total inorganic carbon) value of the sample to
show.
• Measurement Rejection—First Rep: Sets the sample results to not include the first replicate
measurement. Outliers: Sets the sample results to not include the replicate results that are
not within a selected target deviation percentage or ppb. When selected, the user enters a
target deviation percent (or ppb) before a measurement is started. Replicate results are
rejected until the standard deviation or relative standard deviation (the smaller value) is less
than or equal to the target deviation. A minimum of three results are kept if the target
deviation is not achieved.
3. Push Save or Cancel.
Select a report
on page 17 to view and export
English 15