Operation
4.5.3

General information about calibration

4.5.3.1

Clear recorded points

4.5.3.2

Clear a calibration point

24
The value calculated in the laboratory can be corrected at any time by overwriting it.
Points that have been saved under MEMORY can be reset and cleared at any time.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
A mark "<<" is shown after the recorded point or points for approximately 3 seconds.
6. Select MEMORY and confirm.
7. Select CLEAR MEMORY and confirm.
The sensor will continue working with the old calibration values.
8. Go back to the MAIN MENU or the measurement mode display.
An individual calibration point can be cleared at any time by entering the value 0.0 for the
concentration.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
6. Select the POINT to be cleared and confirm.
7. Enter the value 0 and confirm.
8. Go back to the MAIN MENU or the measurement mode display.