4.5.1

Calibration of the TURBIDITY (TRB) parameter

4.5.1.1

Select the TURBIDITY (TRB) parameter

4.5.1.2

FACTOR

4.5.1.3

OFFSET

Installation conditions in the pipes can cause interfering ground reflection when
measuring turbidity, which in turn may cause the zero point to shift. Compensate for this
effect with an offset correction (section
unrelated to the factors described above between the measurement values shown and the
laboratory results, the slope of the calibration curve can be adjusted using a factor (refer to
section
4.5.1 Calibration of the TURBIDITY (TRB)
At least a 1-point calibration must be carried out for a solid matter measurement. In
difficult application conditions, a 2-point or 3-point calibration may be necessary (refer to
section
4.5.2 Calibration of the SOLID (TS)
Before the sensor can be calibrated to the TURBIDITY (TRB) parameter, the parameter
must be selected.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CONFIGURE and confirm.
5. Select PARAMETERS and confirm.
6. Select the TRB parameter and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select FACTOR and confirm.
6. Set the desired factor and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Press OFFSET and confirm.
6. Set the desired offset and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
4.5.1.3
OFFSET). If there are deviations that are
parameter).
parameter).
Operation
21